IEEE Trans Med Imaging. 1996;15(3):260-7. doi: 10.1109/42.500135.
The authors present a novel technique for large-field digital mammography. The instrument uses a mosaic of electronic digital imaging [charge coupled device (CCD)] arrays, novel area scanning, and a radiation exposure and scatter reducing mechanism. The imaging arrays are mounted on a carrier platform in a checker-board pattern mosaic. To fill in the gaps between array-active areas the platform Is repositioned three times and four X-ray exposures are made. The multiple image areas are then recombined by a digital computer to produce a composite image of the entire region. To reduce X-ray scatter and exposure, a lead aperture plate is interposed between X-ray source and patient. The aperture plate has a mosaic of square holes in alignment with the imaging array pattern and the plate is repositioned in synchronism with the carrier platform. The authors discuss proof-of-concept testing demonstrating technical feasibility of their approach. The instrument should be suitable for incorporation into standard mammography units. Unique features of the new technique are: large field coverage (18x24 cm); high spatial resolution (14-17 lp/mm); scatter rejection; and excellent contrast characteristics and lesion detectability under clinical conditions.
作者提出了一种用于大视场数字乳腺摄影的新方法。该仪器使用电子数字成像[电荷耦合器件 (CCD)]阵列的马赛克、新型区域扫描以及辐射暴露和散射减少机制。成像阵列安装在载体平台上,呈棋盘状马赛克图案。为了填补阵列活动区域之间的空白,平台被重新定位三次,进行四次 X 射线曝光。然后,通过数字计算机将多个图像区域重新组合,生成整个区域的复合图像。为了减少 X 射线散射和曝光,在 X 射线源和患者之间插入铅孔径板。孔径板具有与成像阵列图案对齐的正方形孔的马赛克,并且板与载体平台同步重新定位。作者讨论了概念验证测试,证明了他们方法的技术可行性。该仪器应该适合集成到标准乳腺摄影装置中。新技术的独特特点是:大视场覆盖(18x24 厘米);高空间分辨率(14-17 lp/mm);散射拒绝;以及在临床条件下出色的对比度特性和病变检测能力。