A biomechanical and histological analysis of standard versus hydroxyapatite-coated pins for external fixation.

作者信息

Moroni Antonio, Cadossi Matteo, Romagnoli Matteo, Faldini Cesare, Giannini Sandro

机构信息

Rizzoli Orthopaedic Institute, University of Bologna, Bologna, Italy.

出版信息

J Biomed Mater Res B Appl Biomater. 2008 Aug;86(2):417-21. doi: 10.1002/jbm.b.31036.

Abstract

This sheep study was designed to make a comparative evaluation of two external fixation pin types each with and without hydroxyapatite (HA) coating. The two pins had different taper, pitch, and self drilling capabilities. Twenty Orthofix standard, self-tapping pins (group A), 20 Orthofix HA-coated, self-tapping pins (group B), 20 X-caliber, self-drilling, self-tapping pins (group C), and 20 X-caliber HA-coated, self-drilling, self-tapping pins (group D) were selected. Four pins were implanted in the right femurs of 20 adult sheep that were euthanized at 6 weeks. Mean pin insertion torque was 2745 +/- 822 Nmm in group A, 2726 +/- 784 Nmm in group B, 2818 +/- 552 Nmm in group C, and 2657 +/- 732 Nmm in group D (ns). Mean pin extraction torque was 1567 +/- 541 Nmm in group A, 2524 +/- 838 Nmm in group B, 1650 +/- 650 Nmm in group C, and 2517 +/- 726 Nmm in group D. HA-coated pins (group B and D) had a significantly greater mean pin extraction torque compared to similar uncoated pins (group A and C) (p < 0.0005). Histological analysis showed good osteointegration of the two coated pin types. This study shows that HA-coating is more important for optimal pin fixation than the particular combination of design parameters used in each pin type.

摘要

这项绵羊研究旨在对两种分别有无羟基磷灰石(HA)涂层的外固定针进行比较评估。这两种针具有不同的锥度、螺距和自钻能力。选取了20根Orthofix标准自攻针(A组)、20根Orthofix HA涂层自攻针(B组)、20根X-caliber自钻自攻针(C组)和20根X-caliber HA涂层自钻自攻针(D组)。将4根针植入20只成年绵羊的右股骨,6周后对绵羊实施安乐死。A组的平均进针扭矩为2745±822 Nmm,B组为2726±784 Nmm,C组为2818±552 Nmm,D组为2657±732 Nmm(无显著差异)。A组的平均拔针扭矩为1567±541 Nmm,B组为2524±838 Nmm,C组为1650±650 Nmm,D组为2517±726 Nmm。与未涂层的类似针(A组和C组)相比,HA涂层针(B组和D组)的平均拔针扭矩显著更大(p<0.0005)。组织学分析显示两种涂层针类型均有良好的骨整合。这项研究表明,对于最佳的针固定,HA涂层比每种针类型所使用的设计参数的特定组合更为重要。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验