J Electromyogr Kinesiol. 2009 Aug;19(4):704-9. doi: 10.1016/j.jelekin.2008.01.006. Epub 2008 Mar 7.
Flexible goniometers are useful for direct movement measurements. Crosstalk due to rotation between the endblocks is well known. However, even without any rotation, some crosstalk can occur. The objective of this study was to elucidate the effect of, and compensate for, the inherent crosstalk in biaxial goniometers, with specific relevance for applications with one dominating movement direction. Six biaxial goniometers (M110, Biometrics Ltd., Gwent, UK) were evaluated. A precision jig, for simulating pure flexion/extension angles, was constructed. Each sensor produced a consistent and specific crosstalk pattern, when tested over a +/-100 degrees range of motion. A procedure for correction for the inherent crosstalk of individual goniometer, based on polynomial adjust, is presented. The method for compensation, which reduced the root mean square error from, on average for the six goniometers, 3.7 degrees (range 1.8-10.1 degrees) to 0.35 degrees (0.12-0.55 degrees), might be required for obtaining valid goniometer measurements, e.g. of valgus/varus of the knee during gait flexion/extension movements.
柔性测角仪对于直接测量运动很有用。由于端块之间的旋转而产生的串扰是众所周知的。然而,即使没有任何旋转,也可能会出现一些串扰。本研究的目的是阐明双轴测角仪中固有串扰的影响并进行补偿,这对于具有一个主导运动方向的应用具有特殊意义。对六个双轴测角仪(M110,Biometrics有限公司,英国格温特郡)进行了评估。构建了一个用于模拟纯屈伸角度的精密夹具。当在+/-100度的运动范围内进行测试时,每个传感器都产生了一致且特定的串扰模式。提出了一种基于多项式调整来校正单个测角仪固有串扰的程序。为了获得有效的测角仪测量结果,例如在步态屈伸运动期间膝关节的外翻/内翻测量结果,可能需要这种补偿方法,该方法将六个测角仪的均方根误差平均从3.7度(范围为1.8 - 10.1度)降低到0.35度(0.12 - 0.55度)。