Dentkos Terence R, Berzins David W
Department of Graduate Endodontics, Marquette University, Milwaukee, Wisconsin 53233, USA.
J Endod. 2008 Jul;34(7):863-5. doi: 10.1016/j.joen.2008.04.007.
The purpose of this in vitro study was to evaluate the cutting efficiency of 4 orthograde ultrasonic tips and describe a nonstatic cutting model. Forty molars were extracted and decoronated. Each group contained 10 tips. The samples were fixed to a double pan balance that applied a constant vertical force of 40 g to the ultrasonic tips. The ultrasonic handpiece was mounted to a cyclic axial fatigue testing apparatus to generate a 3-mm horizontal movement of the ultrasonic tip during the cutting procedure. Three cutting cycles of 30 seconds were tested for each power setting of 3, 5, and 7. The weight of the samples was measured before and after the 3 cycles. The cutting efficiency was determined by the weight change caused by dentin removal. The Buc-1 ultrasonic tip had a significantly greater cutting efficiency than the other 3 tips at each of the 3 power settings (P < .05).
本体外研究的目的是评估4种直上型超声刀头的切割效率,并描述一种非静态切割模型。提取并去除40颗磨牙的牙冠。每组包含10个刀头。将样本固定在一个双盘天平上,该天平对超声刀头施加40 g的恒定垂直力。将超声手机安装在循环轴向疲劳测试装置上,以便在切割过程中使超声刀头产生3 mm的水平移动。对3、5和7的每个功率设置测试3个30秒的切割周期。在3个周期前后测量样本的重量。切割效率由牙本质去除引起的重量变化确定。在3个功率设置中的每一个设置下,Buc-1超声刀头的切割效率均显著高于其他3个刀头(P < .05)。