Ogura Akio, Hayakawa Katsumi, Miyati Tosiaki, Maeda Fumie
Department of Radiology, Kyoto City Hospital, 1-2 Higashitaka-cho, Mibu, Nakagyo-ku, Kyoto, Japan.
Acad Radiol. 2008 Jul;15(7):867-72. doi: 10.1016/j.acra.2007.12.020.
The development of parallel magnetic resonance imaging has resulted in the frequent use of diffusion-weighted imaging (DWI) in clinical medicine, which usually involves the use of contrast medium. However, gadolinium (Gd) contrast medium may have some effect on DWI and the apparent diffusion coefficient (ADC). The present study was performed to determine whether the magnetic susceptibility of contrast medium alters the DWI signal and the value of ADC in some imaging techniques.
Nonfat suppression DWI, short-time inversion recovery (STIR) combination, and chemical shift selective (CHESS) combination DWI were performed to examine 10 phantoms with gadolinium-meglumine gadopentetate (Gd-DTPA) dissolved at concentrations from 0.0005 to 0.1 mmol in physiologic saline as a contrast medium. The average pixel value and ADC of each method were determined.
ADC showed no differences between before and after treatment with contrast medium for all imaging techniques with Gd considered distributed over the whole tumor. The signal intensity did not change on nonfat suppression or CHESS combination DWI, but deteriorated on STIR.
ADC was not influenced by the magnetic susceptibility of contrast medium. In addition, it was suggested that the ability of tumor detection may be reduced if STIR is used as fat suppression.
并行磁共振成像技术的发展使得扩散加权成像(DWI)在临床医学中频繁使用,而这通常需要使用造影剂。然而,钆(Gd)造影剂可能会对DWI和表观扩散系数(ADC)产生一定影响。本研究旨在确定造影剂的磁化率是否会在某些成像技术中改变DWI信号和ADC值。
使用钆喷酸葡胺(Gd-DTPA)作为造影剂,以0.0005至0.1 mmol的浓度溶解于生理盐水中,对10个模型进行非脂肪抑制DWI、短时反转恢复(STIR)组合以及化学位移选择性(CHESS)组合DWI检查。确定每种方法的平均像素值和ADC。
对于所有成像技术,在考虑Gd分布于整个肿瘤的情况下,造影剂处理前后ADC无差异。非脂肪抑制或CHESS组合DWI上信号强度未改变,但在STIR上信号强度变差。
ADC不受造影剂磁化率的影响。此外,研究表明,如果使用STIR进行脂肪抑制,肿瘤检测能力可能会降低。