使用光子多叶准直器对完整乳房进行调强电子放疗的剂量验证。

Dosimetric verification of modulated electron radiotherapy delivered using a photon multileaf collimator for intact breasts.

作者信息

Jin L, Ma C-M, Fan J, Eldib A, Price R A, Chen L, Wang L, Chi Z, Xu Q, Sherif M, Li J S

机构信息

Department of Radiation Oncology, Fox Chase Cancer Center, Philadelphia, PA 19111, USA.

出版信息

Phys Med Biol. 2008 Nov 7;53(21):6009-25. doi: 10.1088/0031-9155/53/21/008. Epub 2008 Oct 3.

Abstract

Modulated electron radiotherapy (MERT) may potentially be an effective modality for the treatment of shallow tumors, but dose calculation accuracy and delivery efficiency challenges remain. The purpose of this work is to investigate the dose accuracy of MERT delivery using a photon multileaf collimator (pMLC) on a Siemens Primus accelerator. A Monte Carlo (MC)-based inverse treatment planning system was developed for the 3D treatment planning process. Phase space data of 6, 9, 12 and 15 MeV electron beams were commissioned and used as the input source for MC dose calculations. A treatment plan was performed based on the 3D CT data of a heterogeneous 'breast phantom' that mimics a breast cancer patient, and delivered with 22 segments, each associated with a particular energy and Monitor Unit value. Film and ion chamber dosimetry was carefully performed for the conversion from measurement reading to dose, and the results were employed for plan verification using the heterogeneous breast phantom and a solid water phantom. Dose comparisons between measurements and calculations showed agreement within 2% or 1 mm. We conclude that our in-house MC treatment planning system is capable of performing treatment planning and accurate dose calculations for MERT using the pMLC to deliver radiation therapy to the intact breast.

摘要

调制电子放疗(MERT)可能是治疗浅表肿瘤的一种有效方式,但剂量计算准确性和输送效率方面仍存在挑战。本研究的目的是在西门子Primus加速器上,使用光子多叶准直器(pMLC)研究MERT输送的剂量准确性。基于蒙特卡洛(MC)的逆向治疗计划系统被开发用于三维治疗计划过程。对6、9、12和15兆电子伏电子束的相空间数据进行了调试,并用作MC剂量计算的输入源。基于模拟乳腺癌患者的非均匀“乳腺体模”的三维CT数据执行了一个治疗计划,并以22个射野进行输送,每个射野与特定能量和监测单位值相关联。为了从测量读数转换为剂量,仔细进行了胶片和电离室剂量测定,并将结果用于使用非均匀乳腺体模和固体水模体进行计划验证。测量值与计算值之间的剂量比较显示在2%或1毫米范围内一致。我们得出结论,我们内部的MC治疗计划系统能够使用pMLC为MERT进行治疗计划和准确的剂量计算,从而向完整乳腺输送放射治疗。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验