Marmor M F, Fulton A B, Holder G E, Miyake Y, Brigell M, Bach M
Department of Ophthalmology, Stanford University School of Medicine, Stanford, CA, USA.
Doc Ophthalmol. 2009 Feb;118(1):69-77. doi: 10.1007/s10633-008-9155-4. Epub 2008 Nov 22.
This document, from the International Society for Clinical Electrophysiology of Vision (ISCEV), presents an updated and revised ISCEV Standard for clinical electroretinography (ERG). The parameters for flash stimulation and background adaptation have been tightened, and responses renamed to indicate the flash strength (in cd x s x m(-2)). The ISCEV Standard specifies five responses: (1) Dark-adapted 0.01 ERG (rod response); (2) Dark-adapted 3.0 ERG (combined rod-cone response); (3) Dark-adapted 3.0 oscillatory potentials; (4) Light-adapted 3.0 ERG (cone response); (5) Light-adapted 3.0 flicker (30 Hz flicker). An additional Dark-adapted 10.0 ERG or Dark-adapted 30.0 ERG response is recommended.
这份来自国际视觉临床电生理学会(ISCEV)的文件,介绍了更新和修订后的临床视网膜电图(ERG)的ISCEV标准。闪光刺激和背景适应的参数已收紧,反应名称重新命名以表明闪光强度(单位为cd x s x m(-2))。ISCEV标准规定了五种反应:(1)暗适应0.01 ERG(视杆细胞反应);(2)暗适应3.0 ERG(视杆-视锥细胞联合反应);(3)暗适应3.0振荡电位;(4)明适应3.0 ERG(视锥细胞反应);(5)明适应3.0闪烁(30Hz闪烁)。建议增加暗适应10.0 ERG或暗适应30.0 ERG反应。