Suppr超能文献

用于肢体磁共振成像系统的局部电磁屏蔽的研制。

Development of a local electromagnetic shielding for an extremity magnetic resonance imaging system.

作者信息

Handa Shinya, Haishi Tomoyuki, Kose Katsumi

机构信息

Institute of Applied Physics, University of Tsukuba, Tsukuba, Ibaraki, Japan.

出版信息

Rev Sci Instrum. 2008 Nov;79(11):113706. doi: 10.1063/1.3030856.

Abstract

A local radio frequency (rf) shielding consisting of a Cu plate and an LC balun circuit has been developed for a compact magnetic resonance imaging (MRI) system with a 0.3 T permanent magnet. Performance of the local rf shielding was evaluated using an artificial external noise source irradiating a human subject whose hand was inserted into the rf coil of the MRI system. Power spectra of the rf signal detected through the rf coil demonstrated that the local rf shield achieved 30.1 dB external noise suppression. With the local rf shielding, a MRI of the subject's hand was performed using a three-dimensional gradient-echo sequence. Anatomical structures of the subject's hand were clearly visualized. It was concluded that the local rf shielding could be used for the compact MRI system instead of a rf shielded room.

摘要

已为具有0.3 T永磁体的紧凑型磁共振成像(MRI)系统开发了一种由铜板和LC巴伦电路组成的局部射频(rf)屏蔽装置。使用人工外部噪声源照射手部插入MRI系统射频线圈的人体受试者,对局部rf屏蔽的性能进行了评估。通过射频线圈检测到的射频信号的功率谱表明,局部rf屏蔽实现了30.1 dB的外部噪声抑制。使用局部rf屏蔽,采用三维梯度回波序列对受试者的手部进行了MRI检查。受试者手部的解剖结构清晰可见。得出的结论是,局部rf屏蔽可用于紧凑型MRI系统,而无需使用rf屏蔽室。