Ringel Florian, Ingerl Dominik, Ott Stephanie, Meyer Bernhard
Department of Neurosurgery, Technical University of Munich, Munich, Germany.
Neurosurgery. 2009 May;64(5 Suppl 2):365-71; discussion 371-3. doi: 10.1227/01.NEU.0000341532.15867.1C.
VarioGuide (BrainLAB AG, Feldkirchen, Germany) is a new system for frameless image-guided stereotaxy. In the present study, we aimed to assess target point accuracy in a laboratory setting and the clinical feasibility of the system.
Using the phantom of our frame-based stereotactic system (Riechert-Mundinger; Inomed Medizintechnik GmbH, Teningen, Germany), target points were approached from different angles with the frameless system. Target point deviation in the x, y, and z planes was assessed. Furthermore, patients harboring intracranial lesions were diagnostically biopsied using VarioGuide.
Phantom-based accuracy measurements yielded a mean target point deviation of 0.7 mm. Between February 2007 and April 2008, 27 patients were diagnostically biopsied. Lesion volumes ranged from 0.2 to 117.6 cm3, trajectory length ranged from 25.3 to 64.1 mm, and the diagnostic yield was 93%.
Concluding from the phantom measurements with ideal image-object registration, assumed spherical lesions with a volume of 0.524 cm can be biopsied with 100% target localization. Early clinical data revealed VarioGuide to be safe and accurate for lesions of 0.2 cm3 and larger. Thereby, the system seems feasible for the biopsy of most intracranial lesions.
VarioGuide(德国费尔德kirchen的BrainLAB AG公司)是一种新型的无框架图像引导立体定向系统。在本研究中,我们旨在评估该系统在实验室环境中的靶点定位准确性及其临床可行性。
使用我们基于框架的立体定向系统(Riechert-Mundinger;德国特宁根的Inomed Medizintechnik GmbH公司)的模型,用无框架系统从不同角度接近靶点。评估靶点在x、y和z平面的偏差。此外,使用VarioGuide对患有颅内病变的患者进行诊断性活检。
基于模型的准确性测量得出靶点平均偏差为0.7毫米。在2007年2月至2008年4月期间,对27例患者进行了诊断性活检。病变体积范围为0.2至117.6立方厘米,轨迹长度范围为25.3至64.1毫米,诊断成功率为93%。
根据理想图像-物体配准的模型测量结果,体积为0.524立方厘米的假定球形病变活检时靶点定位准确率可达100%。早期临床数据显示,VarioGuide对于0.2立方厘米及以上的病变安全且准确。因此,该系统对于大多数颅内病变的活检似乎是可行的。