Measurement of relative output factors for the 8 and 4 mm collimators of Leksell Gamma Knife Perfexion by film dosimetry.

作者信息

Novotny Josef, Bhatnagar Jagdish P, Quader Mubina A, Bednarz Greg, Lunsford L Dade, Huq M Saiful

机构信息

Department of Radiation Oncology, University of Pittsburgh Cancer Institute, Pittsburgh, Pennsylvania 15213, USA.

出版信息

Med Phys. 2009 May;36(5):1768-74. doi: 10.1118/1.3113904.

Abstract

Three types of films, Kodak EDR2, Gafchromic EBT, and Gafchromic MD-V2-55, were used to measure relative output factors of 4 and 8 mm collimators of the Leksell Gamma Knife Perfexion. The optical density to dose calibration curve for each of the film types was obtained by exposing the films to a range of known doses. Ten data points were acquired for each of the calibration curves in the dose ranges from 0 to 4 Gy, 0 to 8 Gy, and 0 to 80 Gy for Kodak EDR2, Gafchromic EBT, and Gafchromic MD-V2-55 films, respectively. For the measurement of relative output factors, five films of each film type were exposed to a known dose. All films were scanned using EPSON EXPRESSION 10000 XL scanner with 200 dpi resolution in 16 bit gray scale for EDR2 film and 48 bit color scale for Gafchromic films. The scanned images were imported in the red channel for both Gafchromic films. The background corrections from an unexposed film were applied to all films. The output factors obtained from film measurements were in a close agreement both with the Monte Carlo calculated values of 0.924 and 0.805 for 8 and 4 mm collimators, respectively. These values are provided by the vendor and used as default values in the vendor's treatment planning system. The largest differences were noted for the Kodak EDR 2 films (-2.1% and -4.5% for 8 and 4 mm collimators, respectively). The best agreement observed was for EBT Gafchromic film (-0.8% and +0.6% differences for 8 and 4 mm collimators, respectively). Based on the present values, no changes in the default relative output factor values were made in the treatment planning system.

摘要

使用三种类型的胶片,即柯达EDR2、Gafchromic EBT和Gafchromic MD-V2-55,来测量Leksell伽玛刀Perfexion的4毫米和8毫米准直器的相对输出因子。通过将胶片暴露于一系列已知剂量来获得每种胶片类型的光密度与剂量校准曲线。对于柯达EDR2、Gafchromic EBT和Gafchromic MD-V2-55胶片,分别在0至4 Gy、0至8 Gy和0至80 Gy的剂量范围内为每条校准曲线获取10个数据点。为了测量相对输出因子,每种胶片类型的五张胶片被暴露于已知剂量。所有胶片均使用爱普生EXPRESSION 10000 XL扫描仪进行扫描,对于EDR2胶片,扫描分辨率为200 dpi,灰度为16位;对于Gafchromic胶片,扫描分辨率为200 dpi,色彩为48位。两种Gafchromic胶片的扫描图像均导入红色通道。对所有胶片应用未曝光胶片的背景校正。从胶片测量获得的输出因子与蒙特卡洛计算值非常吻合,8毫米和4毫米准直器的计算值分别为0.924和0.805。这些值由供应商提供,并在供应商的治疗计划系统中用作默认值。柯达EDR 2胶片的差异最大(8毫米和4毫米准直器分别为-2.1%和-4.5%)。观察到的最佳吻合情况是EBT Gafchromic胶片(8毫米和4毫米准直器的差异分别为-0.8%和+0.6%)。基于当前值,治疗计划系统中未对默认相对输出因子值进行更改。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验