Suppr超能文献

使用Myomed 932对盆底肌肉力量进行阴道测量的重测信度及同一评估者内部信度。

Test-retest intra-rater reliability of vaginal measurement of pelvic floor muscle strength using Myomed 932.

作者信息

Sigurdardottir Thorgerdur, Steingrimsdottir Thora, Arnason Arni, Bø Kari

机构信息

Department of Physiotherapy, University of Iceland, Reykjavik, Iceland.

出版信息

Acta Obstet Gynecol Scand. 2009;88(8):939-43. doi: 10.1080/00016340903093567.

Abstract

OBJECTIVES

To investigate test-retest intra-rater reliability of measurements of pelvic floor muscle (PFM) strength using Myomed 932 (Enraf-Nonius, Delft, the Netherlands).

DESIGN

Test-retest intra-rater reliability test with a 2-7 day interval.

SETTING

Physiotherapy outpatient clinic in association with the Department of Physiotherapy, University of Iceland.

POPULATION

Twenty healthy female volunteers in Reykjavik metropolitan area, Iceland, with a mean age of 43.8 (SD 9.8) years, mean body mass index (BMI) 26.2 (SD 5.3), and mean parity 2.1 children (SD 1.0).

MAIN OUTCOME MEASURES

PFM strength during two attempts of maximal voluntary contraction (MVC) measured as vaginal squeeze pressure in hectoPascals (hPa).

RESULTS

Mean MVC was 34.45 hPa (95% CI: 26.07-42.83) on Day 1 and 34.05 hPa (95% CI: 26.66-41.44) on Day 2. The measurements ranged from 5-74 hPa interindividually on Day 1 and 5-62 hPa on Day 2. Intraclass Correlation Coefficient (ICC) was 0.97; p < 0.001 and coefficient of variation (CV%) was 11.09%. There was no significant difference between the first and second tests (p = 0.74).

CONCLUSION

This test-retest intra-rater reliability study for Myomed 932 showed a very good reliability using ICC. The CV%, however, indicated some degree of intra-individual variation. We suggest that this device can be used both in practice and clinical trials for assessing female PFM strength.

摘要

目的

使用Myomed 932(恩拉夫-诺尼乌斯公司,荷兰代尔夫特)研究盆底肌肉(PFM)力量测量的重测信度(评估者内信度)。

设计

间隔2 - 7天的重测信度(评估者内信度)测试。

地点

与冰岛大学物理治疗系相关的物理治疗门诊。

研究对象

冰岛雷克雅未克市区的20名健康女性志愿者,平均年龄43.8(标准差9.8)岁,平均体重指数(BMI)26.2(标准差5.3),平均生育子女数2.1个(标准差1.0)。

主要观察指标

最大自主收缩(MVC)两次尝试期间的PFM力量,以百帕斯卡(hPa)为单位测量阴道挤压压力。

结果

第1天平均MVC为34.45 hPa(95%可信区间:26.07 - 42.83),第2天为34.05 hPa(95%可信区间:26.66 - 41.44)。个体间测量值第1天范围为5 - 74 hPa,第2天为5 - 62 hPa。组内相关系数(ICC)为0.97;p < 0.001,变异系数(CV%)为11.09%。第一次和第二次测试之间无显著差异(p = 0.74)。

结论

这项针对Myomed 932的重测信度(评估者内信度)研究使用ICC显示出非常好的信度。然而,CV%表明存在一定程度的个体内变异。我们建议该设备可用于实践和临床试验中评估女性PFM力量。