Gregorcic Peter, Pozar Tomaz, Mozina Janez
Faculty of Mechanical Engineering, University of Ljubljana, Askerceva 6, 1000 Ljubljana, Slovenia.
Opt Express. 2009 Aug 31;17(18):16322-31. doi: 10.1364/OE.17.016322.
The influence of quadrature phase shift on the measured displacement error was experimentally investigated using a two-detector polarizing homodyne laser interferometer with a quadrature detection system. Common nonlinearities, including the phase-shift error, were determined and effectively corrected by a robust data-processing algorithm. The measured phase-shift error perfectly agrees with the theoretically determined phase-shift error region. This error is systematic, periodic and severely asymmetrical around the nominal displacement value. The main results presented in this paper can also be used to assess and correct the detector errors of other interferometric and non-interferometric displacement-measuring devices based on phase-quadrature detection.
使用具有正交检测系统的双探测器偏振零差激光干涉仪,通过实验研究了正交相移对测量位移误差的影响。确定了包括相移误差在内的常见非线性,并通过稳健的数据处理算法有效地进行了校正。测量得到的相移误差与理论确定的相移误差区域完全吻合。该误差是系统性的、周期性的,并且在标称位移值附近严重不对称。本文给出的主要结果还可用于评估和校正基于正交相位检测的其他干涉式和非干涉式位移测量装置的探测器误差。