Hans Nidhi, Kapadia Farhad N
P.D. Hinduja Hospital and Medical Research Center, Veer Savarkar Marg, Mahim, Mumbai - 400 016, India.
Indian J Crit Care Med. 2008 Oct;12(4):170-3. doi: 10.4103/0972-5229.45077.
To observe the effects of mobile phone use in the vicinity of medical devices used in a critical care setting.
Electromagnetic interference (EMI) was tested by using two types of mobile phones - GSM and CDMA. Mobile phones were placed at a distance of one foot from three medical devices - syringe pump, mechanical ventilator, and the bedside monitor - in switch off, standby, and talking modes of the phone. Medical devices were observed for any interference caused by the electromagnetic radiations (EMR) from the mobile phones.
Out of the three medical devices that were tested, EMI occurred while using the mobile phone in the vicinity of the syringe pump, in the 'talk mode.' The mean variation observed in the calculated and delivered volume of the syringe pump was 2.66 ml. Mechanical ventilator did not show any specific adverse effects with mobile phone use in the one-foot vicinity. No other adverse effects or unexplained malfunctions or shutdown of the syringe pump, mechanical ventilator, or the bedside monitor was noted during the study period of 36 hours.
EMI from mobile phones have an adverse effect on the medical devices used in critical care setup. They should be used at least one foot away from the diameter of the syringe pump.
观察在重症监护环境中靠近医疗设备使用手机的影响。
使用两种类型的手机——GSM和CDMA测试电磁干扰(EMI)。手机在关机、待机和通话模式下,放置在距离三种医疗设备——注射泵、机械通气机和床边监护仪一英尺远的地方。观察医疗设备是否受到手机电磁辐射(EMR)的任何干扰。
在测试的三种医疗设备中,在注射泵附近使用手机处于“通话模式”时发生了电磁干扰。观察到注射泵计算体积与输送体积的平均变化为2.66毫升。在距离一英尺的范围内使用手机时,机械通气机未显示任何特定的不良影响。在36小时的研究期间,未观察到注射泵、机械通气机或床边监护仪有任何其他不良影响、不明原因的故障或停机。
手机产生的电磁干扰对重症监护环境中使用的医疗设备有不良影响。使用手机时应至少距离注射泵直径一英尺远。