Ilkay Erdoğan, Firat Sakine, Dağli Necati, Ozeke Ozcan
Department of Cardiology, Mesa Hospital, Ankara, Turkey.
Turk Kardiyol Dern Ars. 2010 Jan;38(1):41-3.
A 20-year-old male patient presented with dyspnea and palpitation. An atrial shunt was detected on transthoracic echocardiography. Transesophageal echocardiography (TEE) showed two secundum atrial septal defects (ASD), 13 mm and 15 mm in size, respectively. There was no other congenital heart disease. During right heart catheterization, pulmonary artery pressure was 40 mmHg and the Qp/Qs ratio was 1.9. His coronary arteries were normal. The patient underwent elective percutaneous ASD closure under general anesthesia and continuous TEE monitoring. The distance between the two defects was 16 mm. Two 25-mm PTS sizing balloon catheters were simultaneously inflated, yielding maximum defect diameters of 13 mm and 15 mm. First the smaller defect, then the larger defect were closed with 15 mm and 18 mm Occlutech Figulla ASD occluders, respectively. Total fluoroscopy time was 25 min, and total procedural time was 95 min. No residual shunt was observed. The patient was discharged the next day without any complication.
一名20岁男性患者出现呼吸困难和心悸。经胸超声心动图检查发现心房分流。经食管超声心动图(TEE)显示两个继发孔型房间隔缺损(ASD),大小分别为13毫米和15毫米。无其他先天性心脏病。右心导管检查期间,肺动脉压为40 mmHg,Qp/Qs比值为1.9。其冠状动脉正常。患者在全身麻醉和持续TEE监测下接受择期经皮房间隔缺损封堵术。两个缺损之间的距离为16毫米。同时充盈两根25毫米的PTS sizing球囊导管,使最大缺损直径分别为13毫米和15毫米。先封堵较小的缺损,然后分别用15毫米和18毫米的Occlutech Figulla ASD封堵器封堵较大的缺损。总透视时间为25分钟,总手术时间为95分钟。未观察到残余分流。患者次日出院,无任何并发症。