Vassileva J, Stoyanov D
Department for Radiation Protection at Medical Exposure, National Centre of Radiobiology and Radiation Protection, 132 Kliment Ohridsky blv, 1756 Sofia, Bulgaria.
Radiat Prot Dosimetry. 2010 Apr-May;139(1-3):310-2. doi: 10.1093/rpd/ncq011.
This paper presents the initial experience in performing quality control and patient dose measurements in a cone beam computed tomography (CT) scanner (ILUMA Ultra, IMTEC Imaging, USA) for oral and maxillofacial radiology. The X-ray tube and the generator were tested first, including the kVp accuracy and precision, and the half-value layer (HVL). The following tests specific for panoramic dental systems were also performed: tube output, beam size and beam alignment to the detector. The tests specific for CT included measurements of noise and CT numbers in water and in air, as well as the homogeneity of CT numbers. The most appropriate dose quantity was found to be the air kerma-area product (KAP) measured with a KAP-metre installed at the tube exit. KAP values were found to vary from 110 to 185 microGy m(2) for available adult protocols and to be 54 microGy m(2) for the paediatric protocol. The effective dose calculated with the software PCXMC (STUK, Finland) was 0.05 mSv for children and 0.09-0.16 mSv for adults.
本文介绍了在一台用于口腔颌面放射学的锥形束计算机断层扫描(CT)扫描仪(ILUMA Ultra,美国IMTEC Imaging公司)中进行质量控制和患者剂量测量的初步经验。首先对X射线管和发生器进行了测试,包括千伏峰值(kVp)的准确性和精密度以及半价层(HVL)。还进行了以下针对全景牙科系统的特定测试:管输出、射束大小以及射束与探测器的对准。针对CT的特定测试包括在水和空气中测量噪声和CT值,以及CT值的均匀性。发现最合适的剂量量是通过安装在管出口处的空气比释动能面积乘积(KAP)仪测量的空气比释动能面积乘积(KAP)。对于可用的成人扫描方案,KAP值在110至185微戈瑞·平方米之间变化,而对于儿科扫描方案,KAP值为54微戈瑞·平方米。使用PCXMC软件(芬兰STUK公司)计算的儿童有效剂量为0.05毫希沃特,成人为0.09 - 0.16毫希沃特。