超声生物显微镜与光学相干断层扫描测量人工晶状体眼和无晶状体眼前房中央深度的重复性、可再现性和一致性。

Repeatability, reproducibility, and agreement of central anterior chamber depth measurements in pseudophakic and phakic eyes: optical coherence tomography versus ultrasound biomicroscopy.

机构信息

School of Optometry and Ophthalmology, Eye Hospital, Wenzhou Medical College, Wenzhou, Zhejiang, China.

出版信息

J Cataract Refract Surg. 2010 Jun;36(6):941-6. doi: 10.1016/j.jcrs.2009.12.038.

Abstract

PURPOSE

To evaluate the intraexaminer repeatability, interexaminer reproducibility, and agreement of central anterior chamber depth (ACD) measurements obtained by anterior segment optical coherence tomography (AS-OCT) and ultrasound biomicroscopy (UBM) in pseudophakic eyes and phakic eyes.

SETTING

Eye Hospital, Wenzhou Medical College, Wenzhou, Zhejiang, China.

METHODS

In this prospective study, the central ACD in right pseudophakic eyes and right phakic eyes was measured with the Visante AS-OCT system and the OTI-Scan HF 35-50 UBM system. Intraexaminer repeatability and interexaminer reproducibility were evaluated in a subgroup of eyes from each group. The repeatability and reproducibility of the measurements were evaluated using intraclass correlation coefficients (ICCs) and the Bland-Altman method.

RESULTS

The mean ACD in the 70 pseudophakic eyes was 3.91 mm +/- 0.29 (SD) by AS-OCT and 3.76 +/- 0.33 mm by UBM and in the 70 phakic eyes, 2.97 +/- 0.31 mm and 2.90 +/- 0.32 mm, respectively. The 95% limits of agreement between the 2 methods were -0.32 to 0.62 mm in pseudophakic eyes and -0.11 to 0.25 mm in phakic eyes. The ICC for the repeatability and reproducibility of AS-OCT and UBM measurements was high (range 0.985 to 0.996).

CONCLUSIONS

Both AS-OCT and UBM ACD measurements showed good repeatability and reproducibility. However, the agreement was better in phakic eyes than in pseudophakic eyes. Thus, the measurements should not be used interchangeably in pseudophakic eyes.

摘要

目的

评估眼前节光学相干断层扫描(AS-OCT)和超声生物显微镜(UBM)测量人工晶状体眼和正常眼中央前房深度(ACD)的观察者内重复性、观察者间可重复性和一致性。

设置

中国温州医学院附属眼视光医院。

方法

前瞻性研究中,使用 Visante AS-OCT 系统和 OTI-Scan HF 35-50 UBM 系统分别测量右眼人工晶状体眼和右眼正常眼的中央 ACD。每组的一部分眼评估观察者内重复性和观察者间可重复性。使用组内相关系数(ICCs)和 Bland-Altman 方法评估测量的重复性和可重复性。

结果

70 只人工晶状体眼中,AS-OCT 测量的平均 ACD 为 3.91mm +/- 0.29(SD),UBM 测量的为 3.76 +/- 0.33mm;70 只正常眼中,AS-OCT 测量的平均 ACD 为 2.97 +/- 0.31mm,UBM 测量的为 2.90 +/- 0.32mm。两种方法之间的 95%一致性界限在人工晶状体眼中为-0.32 至 0.62mm,在正常眼中为-0.11 至 0.25mm。AS-OCT 和 UBM 测量的重复性和可重复性的 ICC 较高(范围 0.985 至 0.996)。

结论

AS-OCT 和 UBM ACD 测量均具有良好的重复性和可重复性。然而,在正常眼中的一致性优于在人工晶状体眼中。因此,在人工晶状体眼中,两者的测量结果不能互换使用。