Organic and Biological Mass Spectrometry Group, Chemical Sciences Division, Oak Ridge National Laboratory, Oak Ridge, TN 37831-6131, USA.
Rapid Commun Mass Spectrom. 2010 Jun 30;24(12):1721-9. doi: 10.1002/rcm.4551.
An atmospheric pressure proximal probe thermal desorption sampling method coupled with secondary ionization by electrospray or atmospheric pressure chemical ionization was demonstrated for the mass spectrometric analysis of a diverse set of compounds (dyestuffs, pharmaceuticals, explosives and pesticides) separated on various high-performance thin-layer chromatography plates. Line scans along or through development lanes on the plates were carried out by moving the plate relative to a stationary heated probe positioned close to or just touching the stationary phase surface. Vapors of the compounds thermally desorbed from the surface were drawn into the ionization region of a combined electrospray ionization/atmospheric pressure chemical ionization source where they merged with reagent ions and/or charged droplets from a corona discharge or an electrospray emitter and were ionized. The ionized components were then drawn through the atmospheric pressure sampling orifice into the vacuum region of a triple quadrupole mass spectrometer and detected using full scan, single ion monitoring, or selected reaction monitoring mode. Studies of variable parameters and performance metrics including the proximal probe temperature, gas flow rate into the ionization region, surface scan speed, read-out resolution, detection limits, and surface type are discussed.
大气压近探头热解吸采样方法与电喷雾或大气压化学电离的二次离子化相结合,用于对各种高效薄层色谱板分离的多种化合物(染料、药物、爆炸物和农药)进行质谱分析。通过相对固定加热探头移动板,沿或穿过板上的展开槽进行线扫描,探头靠近或刚好接触固定相表面。从表面热解吸的化合物蒸气被吸入电喷雾电离/大气压化学电离源的电离区域,在那里它们与来自 corona 放电或电喷雾发射器的试剂离子和/或带电液滴合并并被电离。然后,被电离的成分通过大气压采样孔被吸入三重四极杆质谱仪的真空区域,并使用全扫描、单离子监测或选择反应监测模式进行检测。讨论了可变参数和性能指标的研究,包括近探头温度、进入电离区域的气体流速、表面扫描速度、读出分辨率、检测限和表面类型。