Sadeg N, Chuong P H, Claude J R, Hamon M
Laboratoire de Toxicologie, Faculté de Pharmacie, Paris, France.
J Anal Toxicol. 1991 Mar-Apr;15(2):95-7. doi: 10.1093/jat/15.2.95.
In this precise, reversed-phase liquid chromatographic procedure, Cyclosporin A (Cy A) is extracted from 300 microL of whole blood with cyclosporin D as internal standard by liquid-liquid extraction. A 20-microL aliquot of the extract, injected onto a Nucleosil octyl analytical column heated at 72 degrees C, is eluted with a mixture of acetonitrile and 0.01 M phosphate buffer, pH 5.5 at a flow rate of 1 mL/min. Detection is set at 210 nm. The chromatography is complete within 10 min. The detection limit is 25 micrograms/L. Between-run CVs range from 2.3 to 6.2% and recovery is 92.1 +/- 6.3%. The major advantages of our extraction procedure are the rapid clean-up method and the small volume required. This procedure is especially suitable for cyclosporine determination in pediatric transplantations.
在这种精确的反相液相色谱法中,以环孢素D为内标,通过液液萃取从300微升全血中提取环孢素A(Cy A)。取20微升萃取液注入加热至72℃的Nucleosil辛基分析柱,用乙腈和0.01M磷酸盐缓冲液(pH 5.5)的混合物以1毫升/分钟的流速洗脱。检测波长设定为210纳米。10分钟内完成色谱分析。检测限为25微克/升。批间变异系数在2.3%至6.2%之间,回收率为92.1±6.3%。我们的萃取方法的主要优点是快速净化方法和所需体积小。该方法特别适用于儿科移植中环孢素的测定。