Amiridze N S, Darwish R, Griffith G M, Zoarskia G H
University of Maryland School of Medicine, Department of Diagnostic Radiology - Baltimore Veterans Affairs Medical Center,Veterans Affairs Maryland Health Care System, Baltimore, USA -
Interv Neuroradiol. 2008 Jun 30;14(2):165-71. doi: 10.1177/159101990801400207.
Arteriovenous malformations (AVM) of the brain are the most common symptomatic congenital vascular malformation of the central nervous system, with significant associated morbidity and mortality. This study compared the feasibility and efficacy of treating AVMs by embolization with hydrocoils with similar treatment with bare platinum coils, using the swine rete mirabile as a model. A right carotid-jugular fistula was created in ten swine.A microcatheter was positioned into the rete mirabile, which was unilaterally (contralateral to the fistula) embolized with hydrocoils in six and bare platinum coils in four swine. Angiograms were evaluated during and immediately after embolization. Embolization with complete cessation of bloodf low in the unilateral rete mirabile was achieved in all animals treated with hydrocoils. The number of coils needed varied from four to seven (diameter 2-4 mm; mean coil length 22.3 cm). Embolization with platinum coils of similar number (seven) and slightly longer length (mean 37.75 cm) had a minimal effect on blood flow, resulting in occlusion of only small compartments. No immediate complications were noted with either coil. Hydrocoils are more effective in achieving embolization than bare platinum coils. Expansion of the hydrocoil over only a few minutes allows precise placement and stabilization of the coil before detachment. Hydrocoilscan be safely placed into small vessels. This approach may be particularly useful to decrease the flow rate, as a first stage of AVM embolizationin high flow AVMs that contain arteriovenous fistulae.
脑动静脉畸形(AVM)是中枢神经系统最常见的有症状的先天性血管畸形,具有显著的相关发病率和死亡率。本研究以猪奇网为模型,比较了用弹簧圈栓塞治疗AVM与用裸铂金弹簧圈进行类似治疗的可行性和疗效。在10头猪身上建立了右颈动脉 - 颈静脉瘘。将微导管置于奇网内,6头猪用弹簧圈对奇网进行单侧(与瘘对侧)栓塞,4头猪用裸铂金弹簧圈栓塞。在栓塞过程中及栓塞后立即对血管造影进行评估。在用弹簧圈治疗的所有动物中,均实现了单侧奇网血流完全停止的栓塞。所需弹簧圈数量为4至7个(直径2 - 4毫米;平均弹簧圈长度22.3厘米)。用数量相似(7个)但长度稍长(平均37.75厘米)的铂金弹簧圈栓塞对血流影响极小,仅导致小腔隙闭塞。两种弹簧圈均未出现即刻并发症。与裸铂金弹簧圈相比,弹簧圈在实现栓塞方面更有效。弹簧圈在短短几分钟内展开,可在脱离前精确放置并稳定弹簧圈。弹簧圈可安全地放置于小血管中。作为包含动静脉瘘的高流量AVM栓塞治疗的第一阶段以此方法降低流速可能特别有用。