Sinan A A, Thiémélé-Yacé S E, Abouattier-Mansilla E, Vallaeys K, Diemer F
Unité de Formation et Recherche d'Odonto-Stomatologie d'Abidjan, Côte d'Ivoire.
Odontostomatol Trop. 2010 Mar;33(129):23-8.
The Nickel-titanium (Ni-Ti) instruments superelasticity does not does not allow to anticipate rupture.
This study evaluates the effect of electropolishing time on the rupture resistance in flexion fatigue of the endodontics Ni-Ti instruments.
48 HeroShapers 6% 25/100 are divided into 4 groups of surface treatment (without electropolishing, with electropolishing of 70, 80 or 90 seconds). The surface of the instruments is observed under a scanning electron microscope. Each instrument is driving by an inGet control motor at 275 rpm and put on a fatigue bench (stainless steel pipe angled at 60 degrees with r = 10 mm). The number of turn before rupture is recorded by a specific meter. Data were analysed by analysis of variance and a Fischer probable least-square difference test (PLSD) at a significance level of 0.05.
The mean number of rotation to failure in each group of instrument decreases (407 to 355 tours) with the increase of the electropolishing time. There were no statistically significant difference (p = 0, 3508) between the different groups.
Electropolishing from 70 till 90 seconds improves the surface state without modifying the rupture resistance in flexion fatigue of triple helix endodontic instruments with 6% taper.
镍钛(Ni-Ti)器械的超弹性不允许预测其折断情况。
本研究评估电解抛光时间对牙髓治疗镍钛器械弯曲疲劳抗折断性的影响。
将48支6%锥度25/100的HeroShapers器械分为4组表面处理组(未电解抛光、电解抛光70秒、80秒或90秒)。在扫描电子显微镜下观察器械表面。每支器械由inGet控制电机以275转/分钟驱动,并置于疲劳试验台上(与内径为10毫米的不锈钢管成60度角)。通过特定仪表记录折断前的转动次数。数据采用方差分析和费舍尔最小显著差法检验(PLSD)进行分析,显著性水平为0.05。
随着电解抛光时间的增加,每组器械折断时的平均转动次数减少(从407次降至355次)。不同组之间无统计学显著差异(p = 0.3508)。
70至90秒的电解抛光可改善表面状态,而不改变6%锥度三螺旋牙髓治疗器械弯曲疲劳的抗折断性。