Department of Pathobiology/Lerner Research Institute, Cleveland Clinic, Cleveland, OH 44195, USA.
Glycobiology. 2011 Feb;21(2):175-83. doi: 10.1093/glycob/cwq145. Epub 2010 Sep 23.
Hyaluronan (HA) is a glycosaminoglycan found in the extracellular matrix and ranges from several thousand to millions of daltons in size. HA has importance in various pathological conditions and is known to be elevated in several diseases. Three commonly used, commercially available HA enzyme-linked immunosorbent assay (ELISA)-like assays (from Corgenix, Echelon and R&D) were compared on the basis of accuracy, sample variability and ability to measure a range of HA sizes. The Corgenix HA ELISA-like assay displayed the lowest intra-assay variability [coefficient of variation (CV) = 11.7 ± 3.6%], followed by R&D (CV = 12.3 ± 4.6%) and Echelon (CV = 18.9 ± 9.2%). Interassay variability was also lowest for the Corgenix assay (CV = 6.0%), intermediate for the Echelon assay (9.5%) and highest for the R&D assay (CV = 34.1%). The high interassay variability seen for the R&D assay may have been due to the effect of dilution, since the dilution-independent interassay variability was 15.5%. The concentration of the standard HA was overestimated by the Echelon assay by 85% and underestimated by the R&D and Corgenix assays by 34 and 32%, respectively. The Echelon HA ELISA-like assay was the most effective at measuring all sizes of HA tested (2 MDa and 132, 66 and 6.4 kDa), whereas the Corgenix and R&D assays were unable to detect 6.4 kDa HA. These findings suggest that the Echelon HA ELISA-like assay is better suited for size-sensitive HA measurements but has a relatively high variability. The Corgenix and R&D HA ELISA-like assays have low variability and high accuracy but are not suitable for detecting low-molecular-weight HA.
透明质酸(HA)是一种存在于细胞外基质中的糖胺聚糖,分子量从几千到几百万道尔顿不等。HA 在各种病理条件下都很重要,并且在几种疾病中被发现升高。三种常用的、商业上可获得的透明质酸酶联免疫吸附测定(ELISA)样测定法(来自 Corgenix、Echelon 和 R&D)基于准确性、样品变异性和测量一系列 HA 大小的能力进行了比较。Corgenix HA ELISA 样测定法显示出最低的内测定变异性(变异系数(CV)= 11.7 ± 3.6%),其次是 R&D(CV = 12.3 ± 4.6%)和 Echelon(CV = 18.9 ± 9.2%)。Corgenix 测定法的间测定变异性也最低(CV = 6.0%),Echelon 测定法中等(9.5%),R&D 测定法最高(CV = 34.1%)。R&D 测定法的高间测定变异性可能是由于稀释的影响,因为独立于稀释的间测定变异性为 15.5%。Echelon 测定法高估了标准 HA 的浓度,高估了 85%,而 R&D 和 Corgenix 测定法则低估了 34%和 32%。Echelon HA ELISA 样测定法最有效地测量了所有测试的 HA 大小(2 MDa 和 132、66 和 6.4 kDa),而 Corgenix 和 R&D 测定法无法检测 6.4 kDa 的 HA。这些发现表明,Echelon HA ELISA 样测定法更适合于对 HA 大小敏感的测量,但具有相对较高的变异性。Corgenix 和 R&D 的 HA ELISA 样测定法具有低变异性和高准确性,但不适合检测低分子量的 HA。