Moorfields Eye Hospital NHS Foundation Trust, London, United Kingdom.
J Cataract Refract Surg. 2010 Oct;36(10):1665-70. doi: 10.1016/j.jcrs.2010.05.018.
To assess the rotational stability of a single-piece toric hydrophobic acrylic intraocular lens (IOL) during the first 6 postoperative months.
Moorfields Eye Hospital NHS Foundation Trust, London, United Kingdom.
Prospective case series.
Eyes with age-related cataract and corneal astigmatism (1.00 to 3.00 D diopter [D]) were measured by partial coherence interferometry (IOLMaster). Preoperatively, the horizontal axis was marked. Surgical technique included a standardized temporal clear corneal incision, phacoemulsification, and in-the-bag implantation of a toric IOL (AcrySof SN60TT). Rotational stability of the IOL was assessed using retroillumination photographs postoperatively at 1 hour and at 1 and 6 months. Uncorrected (UDVA) and corrected distance visual acuities, residual refractive error, and keratometric and refractive cylinders were measured.
The study enrolled 30 eyes (30 patients). The mean absolute IOL rotation was 2.44 degrees ± 1.84 (SD) at 1 month and 2.66 ± 1.99 degrees at 6 months. The photographic technique had high reproducibility of axis measurement, with consecutive measurements varying by less than 2.0 degrees. The mean UDVA was 0.16 logMAR (range 0.42 to -0.18 logMAR) at 1 month and 0.20 logMAR (range 0.60 to -0.20 logMAR) at 6 months. The residual refractive cylinder by autorefraction was -0.84 D (range -2.00 to 0.00 D) and -0.80 D (range -1.75 to 0.00 D), respectively.
The acrylic toric IOL was rotationally stable within the first 6 months postoperatively. The photographic and axis analysis method to evaluate stability had high reproducibility and detected small changes in rotation.
评估单焦亲水性丙烯酸 toric 人工晶状体(IOL)在术后头 6 个月的旋转稳定性。
英国伦敦 Moorfields Eye Hospital NHS Foundation Trust。
前瞻性病例系列。
通过部分相干干涉测量法(IOLMaster)测量年龄相关性白内障合并角膜散光(1.00 至 3.00 屈光度[D])的眼。术前标记水平轴。手术技术包括标准化颞侧透明角膜切口、超声乳化白内障吸除术和囊袋内植入 toric IOL(AcrySof SN60TT)。术后 1 小时、1 个月和 6 个月,用 retroillumination 照片评估 IOL 的旋转稳定性。测量未矫正(UDVA)和矫正距离视力、残余屈光误差、角膜曲率和屈光散光。
本研究纳入 30 只眼(30 例患者)。1 个月时,IOL 旋转的平均绝对角度为 2.44°±1.84°(标准差),6 个月时为 2.66°±1.99°。轴位测量的摄影技术具有较高的可重复性,连续测量相差小于 2.0°。1 个月时的平均 UDVA 为 0.16 logMAR(范围 0.42 至-0.18 logMAR),6 个月时为 0.20 logMAR(范围 0.60 至-0.20 logMAR)。自动折射测量的残余屈光柱为-0.84 D(范围-2.00 至 0.00 D)和-0.80 D(范围-1.75 至 0.00 D)。
丙烯酸 toric IOL 在术后头 6 个月内旋转稳定。评估稳定性的摄影和轴位分析方法具有较高的重复性,并能检测到微小的旋转变化。