Suppr超能文献

第三代依维莫司洗脱支架的血管反应。

Vascular response to a third generation everolimus-eluting stent.

机构信息

Department of Pediatric Laboratory Medicine, Hospital for Sick Children, Toronto, Ontario, Canada.

出版信息

EuroIntervention. 2010 Sep;6(4):512-9. doi: 10.4244/EIJ30V6I4A85.

Abstract

AIMS

In a non-injured porcine coronary artery model, the aim was to evaluate vascular compatibility of the novel platinum chromium everolimus-eluting PROMUS Element stent as compared to the following control stents: everolimus-eluting PROMUS (XIENCE V), bare metal Element, and polymer-only Element.

METHODS AND RESULTS

Stent pairs (n=228) evenly distributed among the four stent types were implanted in overlap configuration in 79 pigs at a targeted stent-to-artery ratio of 1.1:1. Similar numbers were explanted at each of 7, 30, 90, 180, and 270 days for pathological analysis. No stent-related mortality or morbidity was observed. There were no stent occlusions or strut fractures. The PROMUS Element was more radiopaque than PROMUS (relative densities 9.9 and 9.1, respectively) and demonstrated at all time points vascular compatibility similar to that of the control stents for endothelial cell coverage, inflammatory response, and neointima formation. At 30 days, parastrut fibrin was mild but greater (P<0.0001) for the drug-eluting stents than either for the bare metal or the polymer-only Element; however, by 90 days the fibrin had dissipated.

CONCLUSIONS

In the non-injured porcine coronary artery model, the PROMUS Element demonstrated vascular compatibility equivalent to PROMUS and the bare metal and polymer-only stents.

摘要

目的

在非损伤性猪冠状动脉模型中,评估新型铂铬依维莫司洗脱 Promus Element 支架与以下对照支架的血管相容性:依维莫司洗脱 Promus(Xience V)、裸金属 Element 和仅聚合物 Element。

方法和结果

将支架对(n=228)均匀分布在四种支架类型中,以 1.1:1 的目标支架与动脉比在 79 头猪中重叠植入。在 7、30、90、180 和 270 天的每个时间点,进行病理分析时,分别取出相同数量的支架。未观察到支架相关的死亡率或发病率。没有支架闭塞或支架断裂。Promus Element 比 Promus(相对密度分别为 9.9 和 9.1)更具不透射线性,并且在所有时间点均显示出与对照支架相似的血管相容性,内皮细胞覆盖、炎症反应和新生内膜形成。在 30 天,支架旁纤维蛋白轻度增加(P<0.0001),但药物洗脱支架比裸金属或仅聚合物 Element 更严重;然而,到 90 天,纤维蛋白已经消散。

结论

在非损伤性猪冠状动脉模型中,Promus Element 显示出与 Promus 和裸金属及仅聚合物支架相当的血管相容性。