Colt Edward, Kälvesten Johan, Cook Kenneth, Khramov Nata, Javed Fahad
Department of Endocrinology and Obesity Research Center, St Luke's Roosevelt Hospital and Columbia University, New York, New York.
Int J Body Compos Res. 2010;8(2):41-44.
We have previously shown that surrounding fat causes an increase of up to 21% in bone mineral density (BMD) measured by Lunar 'Intelligent DXA' (iDXA), one of the latest generation dual energy X-ray absorptiometry (DXA) scanners [1]. The purpose of our study was to see if it was possible to avoid this artifact when measuring the BMD of metacarpals II, III, and IV by digital X-ray radiogrammetry (DXR). METHODS: We took X-rays of the bones of a cadaveric left hand which were immobilized in a wooden cradle to preserve an approximate in vivo configuration. The X-rays were digitized into Digital Imaging and Communications in Medicine (DICOM) files which were analyzed using dxr-online (dxr-online, Sectra, Sweden) which uses the same DXR-BMD algorithm previously used by Pronosco X-posure v2 and Sectra Osteoporosis package. The X-rays were repeated four times. We then surrounded the bones with a layer of lard, and again X-rayed four times. This process was repeated with the bones were covered by two layers, and then three layers of lard. RESULTS: The measured DXR-BMD increased by a maximum of 0.44% when the metacarpals were covered by either two or three layers of lard compared with when the metacarpals were not covered by lard. CONCLUSION: The measurement of metacarpal BMD measured by DXR is minimally affected by surrounding lard. The measurement of metacarpal BMD by DXR seems to be a way of avoiding the artifactual change in BMD caused by fat, when it is measured by DXA.
我们之前已经表明,周围脂肪会使通过Lunar“智能双能X线吸收仪”(iDXA)测量的骨矿物质密度(BMD)增加高达21%,iDXA是最新一代双能X线吸收测定法(DXA)扫描仪之一[1]。我们研究的目的是确定在通过数字X线摄影测量法(DXR)测量第二、三、四掌骨的BMD时,是否有可能避免这种伪影。方法:我们对一具尸体左手的骨骼进行X线摄影,该左手固定在木制支架中以保持近似的体内形态。X线照片被数字化为医学数字成像和通信(DICOM)文件,使用dxr-online(dxr-online,Sectra,瑞典)进行分析,该软件使用与Pronosco X-posure v2和Sectra骨质疏松软件包之前使用的相同的DXR-BMD算法。X线摄影重复进行四次。然后我们用一层猪油包裹骨骼,并再次进行四次X线摄影。当骨骼被两层、然后三层猪油覆盖时,重复这个过程。结果:与未被猪油覆盖时相比,当掌骨被两层或三层猪油覆盖时,测得的DXR-BMD最多增加0.44%。结论:DXR测量的掌骨BMD受周围猪油的影响最小。当通过DXA测量时,DXR测量掌骨BMD似乎是一种避免由脂肪引起的BMD伪变化的方法。