Measurement of digital blood flow using the laser Doppler, impedance and strain-gauge methods.

作者信息

Yamamoto Y, Oberg P A

机构信息

Department of Electrical Engineering, Okayama University, Japan.

出版信息

Med Biol Eng Comput. 1990 Mar;28(2):113-8. doi: 10.1007/BF02441765.

Abstract

Digital volume changes and blood flow have been measured with impedance and strain-gauge plethysmography as well as with laser Doppler flowmetry. A good agreement was found between the impedance and strain-gauge flow measurements with a correlation coefficient of 0.905. The laser Doppler method recorded minor changes in finger skin blood flow following changes in posture from 30 cm below heart level to 60 cm above heart level. This result can be explained as a consequence of the limited penetration depth of laser light into the skin or as a sign of autoregulation of skin blood flow. In these experiments the total blood flow to the finger underwent major changes.

摘要

已使用阻抗和应变片体积描记法以及激光多普勒血流仪测量了数字体积变化和血流。阻抗和应变片血流测量之间发现了良好的一致性,相关系数为0.905。激光多普勒方法记录了从心脏水平以下30厘米到心脏水平以上60厘米姿势变化后手指皮肤血流的微小变化。这一结果可以解释为激光穿透皮肤深度有限的结果,或者是皮肤血流自动调节的迹象。在这些实验中,手指的总血流量发生了重大变化。