使用 NEMA NU 2-2007 标准对 PET/CT Discovery-600 进行性能测量。

Performance measurements for the PET/CT Discovery-600 using NEMA NU 2-2007 standards.

机构信息

Medical Physics Unit, San Gerardo Hospital, Via Pergolesi 33, 20052 Monza, Italy.

出版信息

Med Phys. 2011 Feb;38(2):968-74. doi: 10.1118/1.3544655.

Abstract

PURPOSE

The aim of this study was to assess the performance measurements of the new PET/CT system Discovery-600 (D-600, GEMS, Milwaukee, WI).

METHODS

Performance measures were obtained with the National Electrical Manufacturers Association (NEMA) NU 2-2007 procedures.

RESULTS

The transverse (axial) spatial resolution FWHMs were 4.9 (5.6) mm and 5.6 (6.4) mm at 1 and 10 cm off axis, respectively. The sensitivity (average at 0 and 10 cm) was 9.6 cps/kBq. The scatter fraction was 36.6% (low energy threshold: 425 keV). The NEC peak rate (k=1) was 75.2 kcps at 12.9 kBq/cc. The hot contrasts for 10, 13, 17, and 22 mm spheres were 41%, 51%, 62%, and 73% and the cold contrasts for 28 and 37 mm spheres were 68% and 72%.

CONCLUSIONS

The Discovery-600 has good performance for the NEMA NU 2-2007 parameters, particularly in improved sensitivity compared to the scanners of the same Discovery family, D-ST and D-STE.

摘要

目的

本研究旨在评估新的 PET/CT 系统 Discovery-600(D-600,GEMS,密尔沃基,威斯康星州)的性能测量值。

方法

采用美国电器制造商协会(NEMA)NU 2-2007 程序获得性能测量值。

结果

横向(轴向)空间分辨率 FWHM 值分别为 1 和 10 cm 离轴时的 4.9(5.6)mm 和 5.6(6.4)mm。灵敏度(0 和 10 cm 时的平均值)为 9.6 cps/kBq。散射分数为 36.6%(低能阈值:425 keV)。NEC 峰值率(k=1)在 12.9 kBq/cc 时为 75.2 kcps。10、13、17 和 22 mm 球体的热点对比度分别为 41%、51%、62%和 73%,28 和 37 mm 球体的冷点对比度分别为 68%和 72%。

结论

Discovery-600 在 NEMA NU 2-2007 参数方面具有良好的性能,特别是与同一 Discovery 系列的扫描仪 D-ST 和 D-STE 相比,灵敏度得到了提高。