Suppr超能文献

Study on bubble detectors used as personal neutron dosimeters.

作者信息

Zhang Guiying, Ni Bangfa, Li Li, Lv Peng, Tian Weizhi, Wang Zhiqiang, Zhang Chunbao, Luo Hailong, Jiang Shunli, Wang Pingsheng, Huang Donghui, Liu Cunxiong, Xiao Caijin

机构信息

China Institute of Atomic Energy, Beijing 102413, China.

出版信息

Appl Radiat Isot. 2011 Oct;69(10):1453-8. doi: 10.1016/j.apradiso.2011.05.008. Epub 2011 May 27.

Abstract

Neutron bubble detector is so far the only personal neutron dosimeter satisfying the energy response criteria of the International Committee of Radiation Protection 60 (ICRP 60). This paper presents our studies on neutron bubble detectors including the manufacture, the relevant equipments, the basic calibrations and on-site tests for monitoring personal neutron dose. The results of calibrations show that the highest sensitivity so far manufactured by the authors reaches about 4bubbles/μSv, the correlation coefficient of dose response is 0.99, and the in-batch consistency and reproducibility are up to the ISO standards. The results of on-site test show that the in-batch consistency and between-batch consistency are within 15% relative standard uncertainty. The results are directly readable. The detectors are portable, especially suitable for on-site neutron dose monitoring in n-γ mixed radiation fields.