Suppr超能文献

基于 CT 的超声图像引导放射治疗中声速误差校正方法。

A CT based correction method for speed of sound aberration for ultrasound based image guided radiotherapy.

机构信息

Department of Radiation Oncology (MAASTRO), GROW - School for Oncology and Developmental Biology, Maastricht University Medical Center, Maastricht 6201 BN, The Netherlands.

出版信息

Med Phys. 2011 May;38(5):2665-73. doi: 10.1118/1.3583475.

Abstract

PURPOSE

To introduce a correction for speed of sound (SOS) aberrations in three dimensional (3D) ultrasound (US) imaging systems for small but systematic positioning errors in image guided radiotherapy (IGRT) applications. US waves travel at different speeds in different human tissues. Conventional US-based imaging systems assume that SOS is constant in all tissues at 1540 m/s which is an accepted average value for soft tissues. This assumption leads to errors of up to a few millimeters when converting echo times into distances and is a source of systematic errors and image distortion in quantitative US imaging.

METHODS

At simulation, US applications for IGRT provide a computed tomography (CT) image coregistered to a US volume. The CT scan provides the physical density which can be used in an empirical relationship with SOS. This can be used to correct for different SOS in different tissues within the patient. For each US scan line each voxel's axial dimension is rescaled according to the SOS associated to it. This SOS correction method was applied to US scans of a PMMA container filled with either water, a 20% saline water solution or sunflower oil, and the results were compared to the CT. The correction was also applied to an US quality assurance (QA) phantom containing rods with high ultrasound contrast. This phantom was scanned with US through a container filled with the same three liquids. Finally, the algorithm was applied to two clinical cases: a prostate cancer patient and a breast cancer patient.

RESULTS

After the correction was applied to the phantom images, spatial registration between the bottom of the phantom in the US scan and in the CT scan was improved; the difference was reduced from a few millimeters to less than one millimeter for all three different liquids. Reference structures in the QA phantom appeared at more closely corresponding depths in the three cases after the correction, within 0.5 mm. Both clinical cases showed small shifts, up to 3 mm, in the positions of anatomical structures after correction.

CONCLUSIONS

The SOS correction presented increases quantitative accuracy in US imaging which may lead to small but systematic improvements in patient positioning.

摘要

目的

介绍一种用于三维(3D)超声(US)成像系统的声速(SOS)偏差校正方法,用于图像引导放疗(IGRT)应用中的小但系统定位误差。US 波在不同的人体组织中以不同的速度传播。传统的基于 US 的成像系统假设 SOS 在所有组织中都是恒定的,为 1540m/s,这是软组织的公认平均值。当将回波时间转换为距离时,这种假设会导致高达几毫米的误差,并且是定量 US 成像中系统误差和图像失真的来源。

方法

在模拟中,IGRT 的 US 应用提供了与 US 体积配准的计算机断层扫描(CT)图像。CT 扫描提供了可以用于与 SOS 相关的经验关系的物理密度。这可用于校正患者体内不同组织中的不同 SOS。对于每个 US 扫描线,每个体素的轴向尺寸根据与其相关联的 SOS 进行缩放。该 SOS 校正方法应用于充满水、20%盐水溶液或葵花籽油的 PMMA 容器的 US 扫描,并将结果与 CT 进行比较。该校正方法还应用于包含高超声对比棒的 US 质量保证(QA)体模。该体模通过充满相同三种液体的容器用 US 进行扫描。最后,该算法应用于两个临床病例:前列腺癌患者和乳腺癌患者。

结果

校正后应用于体模图像,US 扫描中体模底部与 CT 扫描中的空间配准得到改善;对于所有三种不同的液体,差异从几毫米减小到不到一毫米。校正后,三个病例中的 QA 体模中的参考结构出现在更接近相应深度,误差在 0.5 毫米以内。校正后,两个临床病例中的解剖结构位置都显示出小的移位,最大可达 3 毫米。

结论

提出的 SOS 校正方法提高了 US 成像的定量准确性,这可能导致患者定位的微小但系统的改善。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验