Am J Ophthalmol. 2012 Mar;153(3):515-23. doi: 10.1016/j.ajo.2011.08.031. Epub 2011 Oct 25.
To evaluate the reproducibility of area and volume measurements of retinal pigment epithelium detachments (PEDs) in eyes of patients with age-related macular degeneration using spectral-domain optical coherence tomography imaging and a novel automated, quantitative algorithm.
Prospective study to evaluate a diagnostic technology.
Patients with PEDs associated with age-related macular degeneration underwent spectral-domain optical coherence tomography imaging. Each eye was imaged 5 times, and each scan consisted of a raster pattern comprising 40 000 uniformly spaced A-scans organized as a 200 × 200 A-scan array. Each raster scan covered a retinal area of 6 × 6 mm encompassing the entire PED. A novel algorithm was used to create PED maps that permitted both qualitative and quantitative assessment of PED area and volume. Test-retest standard deviations of PED area and volume measurements were calculated for each eye.
Sixty-three eyes of 58 patients were enrolled in this study. The qualitative appearance and the quantitative measurements of PED area and volume were highly reproducible over the 5 different datasets obtained from each eye. The intraclass correlation coefficient was more than 0.99 for both area and volume measurements obtained using the entire dataset.
A novel algorithm for the qualitative and quantitative assessment of PEDs imaged using spectral-domain optical coherence tomography was shown to be highly reproducible. The ability to measure PED area and volume reliably represents a novel strategy for following disease progression, especially when assessing the response of vascularized PEDs to antiangiogenic therapy.
利用频域光学相干断层扫描成像和一种新的自动定量算法,评估年龄相关性黄斑变性患者的视网膜色素上皮脱离(PED)的面积和体积测量的可重复性。
评估诊断技术的前瞻性研究。
患有与年龄相关性黄斑变性相关的 PED 的患者接受了频域光学相干断层扫描成像。每只眼成像 5 次,每次扫描由一个包含 40000 个均匀间隔的 A 扫描的光栅图案组成,排列成 200×200 的 A 扫描阵列。每个光栅扫描覆盖一个 6×6mm 的视网膜区域,涵盖整个 PED。使用一种新算法创建 PED 图,可对 PED 面积和体积进行定性和定量评估。为每只眼计算 PED 面积和体积测量的测试-重测标准偏差。
本研究纳入了 58 名患者的 63 只眼。PED 面积和体积的定性表现和定量测量在每只眼的 5 个不同数据集上均具有高度可重复性。使用整个数据集获得的面积和体积测量的组内相关系数均大于 0.99。
用于评估使用频域光学相干断层扫描成像的 PED 的定性和定量的新算法被证明具有高度可重复性。可靠地测量 PED 面积和体积代表了一种用于跟踪疾病进展的新策略,特别是在评估血管化 PED 对抗血管生成治疗的反应时。