Suppr超能文献

经导管 Melody 瓣膜在肺动脉和体循环高压血流动力学环境中的短期性能。

Short-term performance of the transcatheter Melody valve in high-pressure hemodynamic environments in the pulmonary and systemic circulations.

机构信息

Department of Cardiology, Children’s Hospital, 300 Longwood Ave., Boston, MA 02115, USA.

出版信息

Circ Cardiovasc Interv. 2011 Dec 1;4(6):615-20. doi: 10.1161/CIRCINTERVENTIONS.111.963389. Epub 2011 Nov 9.

Abstract

BACKGROUND

The Melody valve is approved for percutaneous pulmonary valve replacement in dysfunctional right ventricular outflow tracts. The function of this valve when subjected to high-pressure loads in humans is unknown. The aim of this study was to describe the immediate and short-term results of Melody valves implanted in a high-pressure environment.

METHODS AND RESULTS

Definitions of a high-pressure system were established for Melody valves implanted in the systemic (ie, aortic or mitral position) and pulmonary (ie, right ventricular outflow tract conduit or tricuspid valve annulus) circulations. Implants in these environments were ascertained from databases of the 5 centers that participated in the US Investigational Device Exemption trial. Thirty implants met the inclusion criteria: 23 pulmonary circulation implants (all in the pulmonary position) systemic circulation implants (5 in the native aortic position, 1 in a left ventricle-to-descending aorta conduit, and 1 in the mitral annulus). All pulmonary circulation implants were performed percutaneously in the catheterization laboratory. A hybrid approach (surgical exposure for transcatheter implant) was used for 4 of the aortic implants. There were no procedure-related deaths. Three patients died of nonprocedure- and nonvalve-related causes. At 1 year, freedom from moderate to severe regurgitation was 100%, and freedom from mild regurgitation was 90%. Freedom from moderate to severe stenosis was 86% at 1 year.

CONCLUSIONS

Short-term performance of the Melody valve in high-pressure environments is encouraging, with good valve function in all patients.

摘要

背景

Melody 瓣膜经批准可用于治疗功能失调的右心室流出道的经皮肺动脉瓣置换。该瓣膜在人体中承受高压负荷时的功能尚不清楚。本研究旨在描述在高压环境下植入 Melody 瓣膜的即刻和短期结果。

方法和结果

为植入体循环(即主动脉或二尖瓣位置)和肺循环(即右心室流出道导管或三尖瓣环)的 Melody 瓣膜确立了高压系统的定义。这些环境中的植入物是从参与美国调查设备豁免试验的 5 个中心的数据库中确定的。符合纳入标准的植入物有 30 个:23 个肺循环植入物(均位于肺位置);体循环植入物(5 个位于原生主动脉位置,1 个位于左心室至降主动脉导管,1 个位于二尖瓣环)。所有肺循环植入物均在导管室经皮进行。其中 4 个主动脉植入物采用了杂交方法(经导管植入时进行手术暴露)。无手术相关死亡。3 例患者死于非手术和非瓣膜相关原因。1 年时,中度至重度反流的无复发率为 100%,轻度反流的无复发率为 90%。1 年时,中度至重度狭窄的无复发率为 86%。

结论

在高压环境下,Melody 瓣膜的短期表现令人鼓舞,所有患者的瓣膜功能均良好。