Hernández L, Rivera T, Jimenez Y, Alvarez R, Zeferino J, Vázquez A, Azorín J
Centro de Investigación en Ciencia Aplicada y Tecnología Avanzada-Legaria, IPN. Av. Legaria 694 C.P. 11500 México D.F., Mexico.
Appl Radiat Isot. 2012 Jul;70(7):1307-9. doi: 10.1016/j.apradiso.2011.12.004. Epub 2011 Dec 11.
The aim of the present work was to evaluate the feasibility of the utilization of CaSO4:Dy pellets for X-ray measurements in a general radiology department. Thermoluminescence (TL) response of CaSO4:Dy+PTFE was compared to the TL response of commercial LiF:Mg,Ti (TLD-100) samples. TL pellets were exposed to X-ray beam from X-ray machine CMR for clinical diagnosis purpose. The calibration curve of CaSO4:Dy+PTFE was obtained and it showed a linear response as a function of absorbed dose in air at the studied dose interval. Despite this fact, this material can be used for X-ray beams measurements if appropriate calibration procedures are performed.