评估一种锥形束 CT 伪影减少算法。

Evaluation of a cone beam CT artefact reduction algorithm.

机构信息

Department of Comprehensive Dentistry, University of Texas Health Science Center, San Antonio, TX 78229-3900, USA.

出版信息

Dentomaxillofac Radiol. 2012 Jul;41(5):422-8. doi: 10.1259/dmfr/43691321. Epub 2012 Feb 23.

Abstract

OBJECTIVES

An algorithm and software to reduce metal artefact has been developed recently and is available in the Picasso Master 3D® (VATECH, Hwaseong, Republic of Korea), which under visual assessment produces better quality images than were obtainable previously. The objective of this in vitro study was to investigate whether the metal artefact reduction (MAR) algorithm of the Picasso Master 3D machine reduced the incidence of metal artefacts and increased the contrast-to-noise ratio (CNR) while maintaining the same gray value when there was no metallic body present within the scanned volume.

METHODS

20 scans with a range of 50-90 kVp were acquired, of which 10 had a metallic bead inserted within a phantom. The images obtained were analysed using public domain software (ImageJ; NIH Image, Bethesda, MD). Area histograms were used to evaluate the mean gray level variation of the epoxy resin-based substitute (ERBS) block and a control area. The CNR was calculated.

RESULTS

The MAR algorithm increased the CNR when the metallic bead was present; it enhanced the ERBS gray level independently of the presence of the metallic bead. The image quality also improved as peak tube potential was increased.

CONCLUSION

Improved quality of images and regaining of the control gray values of a phantom were achieved when the MAR algorithm was used in the presence of a metallic bead.

摘要

目的

最近开发了一种用于减少金属伪影的算法和软件,可用于 Picasso Master 3D®(VATECH,韩国华城)。与之前的图像相比,该软件在视觉评估下可生成质量更好的图像。本体外研究的目的是探究 Picasso Master 3D 机的金属伪影减少(MAR)算法是否能在无金属物体存在于扫描体积内时降低金属伪影的发生率并提高对比噪声比(CNR),同时保持相同的灰度值。

方法

采集了 50-90kVp 范围内的 20 个扫描,其中 10 个扫描在体模内插入了金属珠。使用公共域软件(ImageJ;NIH Image,马里兰州贝塞斯达)分析获得的图像。面积直方图用于评估基于环氧树脂的替代物(ERBS)块和对照区域的平均灰度变化。计算 CNR。

结果

MAR 算法在存在金属珠时提高了 CNR;它增强了 ERBS 灰度值,而与金属珠的存在无关。随着管电压峰值的增加,图像质量也得到了改善。

结论

在存在金属珠的情况下使用 MAR 算法可提高图像质量并恢复体模的对照灰度值。

相似文献

1
Evaluation of a cone beam CT artefact reduction algorithm.
Dentomaxillofac Radiol. 2012 Jul;41(5):422-8. doi: 10.1259/dmfr/43691321. Epub 2012 Feb 23.
2
Metal artefact reduction with cone beam CT: an in vitro study.
Dentomaxillofac Radiol. 2012 Mar;41(3):248-53. doi: 10.1259/dmfr/80899839. Epub 2012 Jan 12.
3
Evaluation of the efficacy of a metal artifact reduction algorithm in different cone beam computed tomography scanning parameters.
Oral Surg Oral Med Oral Pathol Oral Radiol. 2017 Jun;123(6):729-734. doi: 10.1016/j.oooo.2017.02.015. Epub 2017 Mar 9.
5
Influence of acquisition parameters on the magnitude of cone beam computed tomography artifacts.
Dentomaxillofac Radiol. 2018 Dec;47(8):20180151. doi: 10.1259/dmfr.20180151. Epub 2018 Jun 27.
6
Are metal artefact reduction algorithms effective to correct cone beam CT artefacts arising from the exomass?
Dentomaxillofac Radiol. 2019 Mar;48(3):20180290. doi: 10.1259/dmfr.20180290. Epub 2019 Jan 28.

引用本文的文献

3
Influence of CBCT parameters on image quality and the diagnosis of vertical root fractures in teeth with metallic posts: an study.
Restor Dent Endod. 2023 Apr 27;48(2):e16. doi: 10.5395/rde.2023.48.e16. eCollection 2023 May.
7
The Impact of Cone-Beam Computed Tomography Exposure Parameters on Peri-Implant Artifacts: A Literature Review.
Cureus. 2022 Mar 10;14(3):e23035. doi: 10.7759/cureus.23035. eCollection 2022 Mar.
9
Influence of CBCT metal artifact reduction on vertical radicular fracture detection.
Imaging Sci Dent. 2021 Mar;51(1):55-62. doi: 10.5624/isd.20200191. Epub 2021 Jan 29.

本文引用的文献

1
Metal artefact reduction with cone beam CT: an in vitro study.
Dentomaxillofac Radiol. 2012 Mar;41(3):248-53. doi: 10.1259/dmfr/80899839. Epub 2012 Jan 12.
3
On cone-beam computed tomography artifacts induced by titanium implants.
Clin Oral Implants Res. 2010 Jan;21(1):100-7. doi: 10.1111/j.1600-0501.2009.01817.x. Epub 2009 Oct 21.
8
Multi-detector CT imaging in the postoperative orthopedic patient with metal hardware.
Eur J Radiol. 2006 Dec;60(3):470-9. doi: 10.1016/j.ejrad.2006.08.008. Epub 2006 Oct 31.
9
Dose reduction in maxillofacial imaging using low dose Cone Beam CT.
Eur J Radiol. 2005 Dec;56(3):413-7. doi: 10.1016/j.ejrad.2005.05.011. Epub 2005 Jun 22.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验