使用钛垫圈对种植体外部连接类型基台螺丝去除扭矩的影响。

Effect of using a titanium washer on the removal torque of an abutment screw in the external connection type of dental implant.

机构信息

Department of Prosthodontics, College of Dentistry, Yonsei University, Seoul, Republic of Korea.

出版信息

Implant Dent. 2012 Apr;21(2):156-9. doi: 10.1097/ID.0b013e31824a02b9.

Abstract

PURPOSE

The aim of this study was to determine the effect of using a washer on the removal torque of an abutment screw in the external connection type of dental implant and abutment complex.

MATERIALS AND METHODS

Four experimental groups with titanium washers of different types (CP titanium or Ti6Al4V titanium alloy) and thicknesses (1.0 or 2.0 mm) included between the abutment and external connection type of implant were investigated. No washer was applied in the control group. The removal torque was measured with a resolution of 0.1 Ncm. One-way analysis of variance (α = 0.05) and Scheffé's post hoc test were used for statistical analysis.

RESULTS

The removal torque was higher in all the groups with a washer than in the control group (α = 0.05), higher with a CP titanium washer than with a titanium alloy washer of the same thickness, and higher with a 2.0-mm-thick washer than with a 1.0-mm-thick washer (α = 0.05).

CONCLUSIONS

Inclusion of a washer between the titanium screw and the abutment seems to increase the removal torque. A 2.0-mm-thick CP titanium washer might provide the largest increase in the removal torque.

摘要

目的

本研究旨在确定在外部连接式牙种植体和基台复合体中使用垫圈对基台螺丝固位扭矩的影响。

材料与方法

研究了 4 个实验组,实验组在基台和外部连接式种植体之间分别使用了不同类型(CP 钛或 Ti6Al4V 钛合金)和厚度(1.0 或 2.0mm)的钛垫圈。在对照组中未使用垫圈。使用分辨率为 0.1Ncm 的扭矩扳手测量固位扭矩。采用单因素方差分析(α=0.05)和Scheffé 事后检验进行统计学分析。

结果

所有使用垫圈的组的固位扭矩均高于对照组(α=0.05),CP 钛垫圈的固位扭矩高于相同厚度的钛合金垫圈,2.0mm 厚的垫圈的固位扭矩高于 1.0mm 厚的垫圈(α=0.05)。

结论

在钛螺丝和基台之间放置垫圈似乎可以增加固位扭矩。2.0mm 厚的 CP 钛垫圈可能会最大程度地增加固位扭矩。