Michotte C, Soon Park Tae, Lee K B, Lee Jong-Man, Lee Sang Han
Bureau International des Poids et Mesures (BIPM), Republic of Korea.
Appl Radiat Isot. 2012 Sep;70(9):1820-4. doi: 10.1016/j.apradiso.2012.02.048. Epub 2012 Feb 28.
In 2010, a comparison of activity measurements of (99m)Tc was carried out at the KRISS using the BIPM's Transfer Instrument of the International Reference System (SIRTI). The transfer instrument and the KRISS coincidence measurement method are briefly described. The degrees of equivalence with the key comparison reference value and between the present KRISS result and the earlier participants in the SIRTI and SIR comparisons have been evaluated. World-wide consistency of activity measurements of (99m)Tc is demonstrated.