Functional MRI Facility, National Institute of Mental Health, National Institutes of Health, Bethesda, Maryland 20892, USA.
Magn Reson Med. 2013 Feb;69(2):402-10. doi: 10.1002/mrm.24266. Epub 2012 Apr 9.
Pseudo-continuous arterial spin labeling (ASL) can provide best signal-to-noise ratio efficiency with a sufficiently long tag at high fields such as 7 T, but it is very sensitive to off-resonance fields at the tagging location. Here, a robust Prescan procedure is demonstrated to estimate the pseudo-continuous ASL radiofrequency phase and gradients parameters required to compensate the off-resonance effects at each vessel location. The Prescan is completed in 1-2 min and is based on acquisition of label/control pair-wise ASL data as a function of the radiofrequency phase increment applied to the pseudo-continuous ASL train. It is shown that this approach can be used to acquire high quality whole-brain pseudo-continuous ASL perfusion data of the human brain at 7 T.
伪连续动脉自旋标记(pseudo-continuous arterial spin labeling,pCASL)在高场(如 7T)下使用足够长的标记可以提供最佳的信噪比效率,但它对标记位置的离频场非常敏感。在这里,展示了一种稳健的预扫描(Prescan)程序,用于估计伪连续 ASL 射频相位和梯度参数,以补偿每个血管位置的离频效应。预扫描在 1-2 分钟内完成,它基于采集标签/控制对 ASL 数据作为应用于伪连续 ASL 序列的射频相位增量的函数。结果表明,该方法可用于在 7T 下获取高质量的人脑全脑 pCASL 灌注数据。