Department of Surgery, University of Colorado School of Medicine, 12631 East 17th Avenue, MS C313, Aurora, CO 80045, USA.
Surg Endosc. 2012 Oct;26(10):2784-8. doi: 10.1007/s00464-012-2279-3. Epub 2012 Apr 27.
This study aimed to quantify the clinical parameters of mono- and bipolar instruments that inhibit pacemaker function. The specific aims were to quantify pacer inhibition resulting from the monopolar instrument by altering the generator power setting, the generator mode, the distance between the active electrode and the pacemaker, and the location of the dispersive electrode.
A transvenous ventricular lead pacemaker overdrive paced the native heart rate of an anesthetized pig. The primary outcome variable was pacer inhibition quantified as the number of beats dropped by the pacemaker during 5 s of monopolar active electrode activation.
Lowering the generator power setting from 60 to 30 W decreased the number of dropped paced events (2.3 ± 1.2 vs 1.6 ± 0.8 beats; p = 0.045). At 30 W of power, use of the cut mode decreased the number of dropped paced beats compared with the coagulation mode (0.6 ± 0.5 vs 1.6 ± 0.8; p = 0.015). At 30 W coagulation, firing the active electrode at different distances from the pacemaker generator (3.75, 7.5, 15, and 30 cm) did not change the number of dropped paced beats (p = 0.314, analysis of variance [ANOVA]). The dispersive electrode was placed in four locations (right/left gluteus, right/left shoulder). More paced beats were dropped when the current vector traveled through the pacemaker/leads than when it did not (1.5 ± 1.0 vs 0.2 ± 0.4; p < 0.001).
Clinical parameters that reduce the inhibition of a pacemaker by monopolar instruments include lowering the generator power setting, using cut (vs coagulation) mode, and locating the dispersive electrode so the current vector does not traverse the pacemaker generator or leads.
本研究旨在量化抑制起搏器功能的单极和双极仪器的临床参数。具体目的是通过改变发生器功率设置、发生器模式、活动电极与起搏器之间的距离以及分散电极的位置来量化单极仪器引起的起搏器抑制。
经静脉心室导联起搏器超速起搏麻醉猪的固有心率。主要结局变量是在 5 秒的单极活动电极激活期间,起搏器抑制的数量,以起搏次数减少来量化。
将发生器功率设置从 60W 降低至 30W 可减少起搏事件(2.3 ± 1.2 次 vs 1.6 ± 0.8 次;p = 0.045)。在 30W 功率下,与凝血模式相比,使用切割模式可减少起搏次数(0.6 ± 0.5 次 vs 1.6 ± 0.8 次;p = 0.015)。在 30W 凝血时,在不同距离(3.75、7.5、15 和 30cm)从起搏器发生器激发活动电极不会改变起搏次数(p = 0.314,方差分析[ANOVA])。分散电极放置在四个位置(右/左臀肌、右/左肩)。当电流向量穿过起搏器/导联时,会比不穿过时多起搏次数(1.5 ± 1.0 次 vs 0.2 ± 0.4 次;p < 0.001)。
降低单极仪器对起搏器抑制的临床参数包括降低发生器功率设置、使用切割(vs 凝血)模式以及定位分散电极,使电流向量不穿过起搏器发生器或导联。