AlGhazali N, Burnside G, Smith R W, Preston A J, Jarad F D
Liverpool University Dental Hospital, Pembroke Place Liverpool.
Eur J Prosthodont Restor Dent. 2011 Dec;19(4):168-74.
Four different shades were used to produce 20 samples of resin-based composite and 20 samples of porcelain to evaluate the performance ability of an intra oral test spectrophotometer compared to a reference spectrophotometer. The absolute colour coordinates CIELAB values measured with both spectrophotometers were significantly different (p < 0.001). However, a high correlation was found (p < 0.001) despite the low concordance noticed. The colour difference deltaE* values calculated between different shades also were significantly different between both spectrophotometers (p < 0.05). Therefore, the Easy Shade can be used in dental practice and dental research with some limitations.