Wust P, Nadobny J, Fähling H, Riess H, Koch K, John W, Felix R
Klinikum Rudolf Virchow, Freie Universität Berlin.
Strahlenther Onkol. 1990 Dec;166(12):822-30.
A new generation of annular-phased-array systems BSD-2000 has been clinically applied in a pilot study. Therapeutic intratumoral temperatures greater than 42 degrees C were obtained in 15/15 sessions with six patients. However, the control of power deposition pattern has to be improved in order to increase the therapeutic gain and to guarantee an efficient therapy. A retrospective analysis of clinical phenomena has been performed by phantom set-ups because the power deposition pattern cannot be determined during therapy. Phantom measurement techniques are outlined, specifically phantom materials and visualization of power distributions. The problem of focus balance and frequency choice is illustrated by self-developed phantoms (liquid crystal sheets, light-emitting-diode-arrays). Especially, the limitation of modeling calculations is demonstrated.
新一代环形相控阵系统BSD - 2000已在一项初步研究中得到临床应用。6名患者的15次治疗中,肿瘤内治疗温度均达到42摄氏度以上。然而,为了提高治疗增益并确保有效治疗,必须改进功率沉积模式的控制。由于在治疗过程中无法确定功率沉积模式,因此通过模型设置对临床现象进行了回顾性分析。概述了模型测量技术,特别是模型材料和功率分布的可视化。通过自行开发的模型(液晶片、发光二极管阵列)说明了焦点平衡和频率选择的问题。特别是,展示了建模计算的局限性。