Validation of a multi-segment spinal model for kinematic analysis and a comparison of different data processing techniques.

作者信息

Needham R, Chockalingam N, Naemi R, Shannon T, Healy A

机构信息

Staffordshire University, Stoke on Trent, UK.

出版信息

Stud Health Technol Inform. 2012;176:151-4.

Abstract

Optoelectronic motion capture technology is a useful tool in the quantitative dynamic assessment of the spine. In a clinical setting this may help gain a further understanding of underlining musculoskeletal pathology. It is therefore important that accurate measurements are made to allow data to be comparable across various investigations. This report outlines a new multi-segment spinal model and its validation. A mechanical model consisting of an upper thoracic (UT), lower thoracic and lumbar segment was developed allowing for range of motion assessment. An electrogoniometer and torsiometer were attached to the model to provide a control measurement. The UT segment was chosen for analysis and static trials were collected at angles ranging from 2-45°. Kinematic data was captured using an optoelectronic motion capture system. Software computed angles corresponded well with the control measure. While highlighting the differences in the estimation of angles between software platforms, this study emphasizes the need for the clear description and understanding of the kinematic model used.