Luo Yong-Heng, Zhong Qian, Ouyang Ping-Bo, Guo Xiao-Jian, Duan Xuan-Chu
Department of Ophthalmology, the Second Xiangya Hospital, Central South University, Changsha 410011, Hunan Province, China.
Int J Ophthalmol. 2012;5(3):329-33. doi: 10.3980/j.issn.2222-3959.2012.03.15. Epub 2012 Jun 18.
To evaluate the repeatability of central corneal thickness (CCT) measurement by entacam, and agreement of CCT measured by Pentacam and ultrasound pachymetry (USP) in Chinese myopia. Thereby investigate the possibility of Pentacam as a substitute for USP in CCT measurement before refractive surgery. The effects of corneal curvature measured by Pentacam on CCT were also evaluated.
One hundred and forty-eight right eyes of 148 individual with myopia were included in this study. Three successive Pentacam CCT measurements followed by 10 successive ultrasound pachymetry were carried out in the 148 eyes. Mean of CCT taken by each device was calculated for comparison. According to the CCT measured by USP, all the 148 eyes were divided into 3 groups: <520µm, 520-560µm, >560µm. For all eyes and each group the CCT obtained by Pentacam and USP were compared. Anterior corneal curvature of the 148 eyes was also adopted for correlation analysis with CCT obtained by ultrasound pachymetry. In addition, CCT measurement using 60 random selected Scheimpflug images was performed by 3 skilled investigators at different time, and this was repeated for 3 times by a forth investigator to assess repeatability of Pentacam CCT measurement using Scheimpflug images.
Intraclass correlation coefficient (ICC) analysis revealed high intraobserver repeatability (ICC=0.994, F=158.60, P<0.001) for CCT measurement by Pentacam. The interobserver (ICC=0.998, F=494.73, P<0.001) and intraobserver (ICC=0.997, F=383.98, P<0.001) repeatability for Pentacam CCT measurements using Scheimpflug images were also excellent. There was high positive correlation between the CCT values measured by Pentacam and ultrasound pachymetry (r=0.963, P<0.001). Bland-altman plots showed that the Pentacam underestimate the CCT by 8.02µm compared with ultrasouond pachymetry. The differences between Pentacam and USP increased as the CCT readings by USP increased (Pentacam vs USP: slope=-0.04, P<0.05). The 95% upper and lower limits of agreement between CCT values obtained from the two devices were +9.33µm and -25.37µm. No significant association could be found between CCT and anterior corneal curvature.
Inter- and intraobserver variability for CCT measurements by Pentacam was considerably below clinically significant levels. CCT of myopia obtained by Scheimpflug camera, Pentacam, were highly correlated to that by ultrasound pachymetry. However, the values obtained are not directly interchangeable between Pentacam and ultrasound pachymetry as the 95% limits of agreement are relatively wide. Pentacam can be a useful instrument for measuring CCT in candidates to refractive surgery in clinic.
评估眼前节分析系统(Pentacam)测量中央角膜厚度(CCT)的可重复性,以及Pentacam测量的CCT与超声角膜测厚仪(USP)在中国近视人群中的一致性。从而探讨在屈光手术前Pentacam替代USP测量CCT的可能性。还评估了Pentacam测量的角膜曲率对CCT的影响。
本研究纳入148例近视患者的148只右眼。对这148只眼进行连续3次Pentacam CCT测量,随后进行连续10次超声角膜测厚。计算每个设备测量的CCT平均值进行比较。根据USP测量的CCT,将148只眼分为3组:<520μm、520 - 560μm、>560μm。比较所有眼以及每组中Pentacam和USP测量的CCT。采用148只眼的前角膜曲率与超声角膜测厚仪测量的CCT进行相关性分析。此外,由3名熟练的研究者在不同时间使用60张随机选择的眼前节图像进行CCT测量,并由第四名研究者重复3次,以评估使用眼前节图像的Pentacam CCT测量的可重复性。
组内相关系数(ICC)分析显示,Pentacam测量CCT具有较高的观察者内可重复性(ICC = 0.994,F = 158.60,P < 0.001)。使用眼前节图像的Pentacam CCT测量的观察者间(ICC = 0.998,F = 494.73,P < 0.001)和观察者内(ICC = 0.997,F = 383.98,P < 0.001)可重复性也很好。Pentacam和超声角膜测厚仪测量的CCT值之间存在高度正相关(r = 0.963,P < 0.001)。Bland - Altman图显示,与超声角膜测厚仪相比,Pentacam低估CCT 8.02μm。随着USP测量的CCT读数增加,Pentacam与USP之间的差异增大(Pentacam与USP:斜率 = -0.04,P < 0.05)。两台设备获得的CCT值之间的95%一致性上限和下限分别为 +9.33μm和 -25.37μm。未发现CCT与前角膜曲率之间存在显著关联。
Pentacam测量CCT的观察者间和观察者内变异性远低于临床显著水平。通过眼前节相机Pentacam获得的近视患者的CCT与超声角膜测厚仪高度相关。然而,由于95%一致性界限相对较宽,Pentacam和超声角膜测厚仪获得的值不能直接互换。Pentacam可作为临床屈光手术候选者测量CCT的有用工具。