Suppr超能文献

膜进样质谱仪用于地下水溶解气体的准连续现场分析。

Membrane inlet mass spectrometer for the quasi-continuous on-site analysis of dissolved gases in groundwater.

机构信息

Department of Water Resources and Drinking Water, Eawag, Swiss Federal Institute of Aquatic Science and Technology, Dübendorf, Switzerland.

出版信息

Environ Sci Technol. 2012 Aug 7;46(15):8288-96. doi: 10.1021/es3004409. Epub 2012 Jul 19.

Abstract

We developed a stand-alone system based on a membrane inlet mass spectrometer (MIMS) for measuring dissolved gas concentrations in groundwater under field conditions. The system permits the concentrations of dissolved gases (He, Ar, Kr, N(2), and O(2)) in groundwater to be determined quasi-continuously (every 12 min) with a precision of better than 4% for He and Kr, and with a precision of 1% for Ar, N(2), and O(2) in air-saturated water. The detection limits are below 3 × 10(-9) cm(3)(STP)(g) for the noble gases and below 400 × 10(-9)cm(3)(STP)(g) for N(2) and O(2). The results of a first deployment of the system in the field indicate that changes in the concentration of Ar that result from diel fluctuations of 3°C in the river water temperature were still able to be resolved in groundwater, although the corresponding temperature signal almost vanished.

摘要

我们开发了一种基于膜进样质谱仪(MIMS)的独立系统,用于在野外条件下测量地下水溶解气体的浓度。该系统允许准连续(每 12 分钟一次)测定地下水溶解气体(He、Ar、Kr、N2 和 O2)的浓度,He 和 Kr 的精度优于 4%,而 Ar、N2 和 O2 在空气饱和水中的精度为 1%。检测限对于稀有气体低于 3×10(-9)cm(3)(STP)(g),对于 N2 和 O2 低于 400×10(-9)cm(3)(STP)(g)。该系统在野外的首次部署结果表明,尽管相应的温度信号几乎消失,但仍能在地下水中分辨出河水温度昼夜波动导致的 Ar 浓度变化。