Frongia Gianluigi, Piancino Maria Grazia, Bracco Pietro
Department of Orthodontics and Gnathology-Masticatory Function, University of Turin, TurinYDental School, Torino, Italy.
J Craniofac Surg. 2012 Jul;23(4):1038-43. doi: 10.1097/SCS.0b013e318252d5e1.
The aim of this research was to analyze the influence of the position of the skull during cone-beam computed tomography (CBCT) scan and if the three-dimensional cephalometric measurements are influenced by skull orientation during CBCT scan.The study consisted of 5 CBCT scanning (KODAK 9500 Cone Beam 3D System unit) in 5 different positions of a dry skull. The data were imported in SIMPLANT OMS Software version 13.0. Fifteen three-dimensional cephalometric measurements were calculated; moreover, the mean, the SD, the maximum/minimum Δ, and the maximum/minimum Δ percentage were calculated. The statistical analysis was performed by an independent-samples t-test to evaluate differences between the 5 scans.No difference was found in all the three-dimensional analysis. Twelve of 15 measurements have a Δ greater than 1.5, and 7 of 15 measurements have a Δ greater than 2. Nine of 15 have a Δ percentage greater than 5%. The preliminary results suggest that the three-dimensional cephalometric analysis is influenced by patient scanning position.
本研究的目的是分析锥形束计算机断层扫描(CBCT)期间颅骨位置的影响,以及CBCT扫描期间颅骨方向是否会影响三维头影测量。该研究包括对一个干燥颅骨在5个不同位置进行5次CBCT扫描(柯达9500锥形束三维系统)。数据被导入SIMPLANT OMS软件13.0版。计算了15项三维头影测量值;此外,还计算了平均值、标准差、最大/最小差值以及最大/最小差值百分比。通过独立样本t检验进行统计分析,以评估5次扫描之间的差异。在所有三维分析中均未发现差异。15项测量中有12项的差值大于1.5,15项测量中有7项的差值大于2。15项中有9项的差值百分比大于5%。初步结果表明,三维头影测量分析受患者扫描位置的影响。