Torrance Clinical Research Institute, Lomita, CA 90717, USA.
Postgrad Med. 2012 Jul;124(4):250-5. doi: 10.3810/pgm.2012.07.2584.
The purpose of this study was to determine the accuracy of noninvasive hemoglobin measurement using pulse CO-oximetry (Pronto® SpHb®, Masimo Corp.) and a commonly used, invasive, point-of-care, automated spectrophotometer (HemoCue 201+®, HemoCue, Inc.), in comparison with hemoglobin measurements obtained from a laboratory hematology analyzer (measuring total blood hemoglobin) in the outpatient setting.
Adult patients presenting to an outpatient research clinic were tested for total blood hemoglobin measurement by 3 methods: noninvasive pulse CO-oximetry (SpHb®), finger-stick blood sample on a point-of-care device, and venous sample on a laboratory hematology analyzer (reference device). Bias and standard deviation (SD) of SpHb® and HemoCue 201+® compared with the values obtained with the laboratory hematology analyzer were calculated and Bland-Altman graphs were generated.
Samples from 152 subjects were assessed (average age, 46 years; 69% female). The bias ± SD compared with the reference method was -0.5 ± 1.0 g/dL for SpHb® and 0.3 ± 1.0 g/dL for HemoCue 201+®. The Bland-Altman plots assessing agreement of the test methods to the reference method had limits of agreement of -2.5 to 1.5 g/dL for SpHb® and -1.7 to 2.3 g/dL for HemoCue 201+®. A noninvasive measurement could not be obtained in 4 subjects after 2 attempts (2.5% failure rate), whereas the HemoCue 201+® measurements were obtained for all subjects.
Noninvasive SpHb® testing had bias and SD similar to those of HemoCue 201+®. Because SpHb® measurement is noninvasive, it may offer additional patient and provider benefits.
本研究旨在确定脉搏 CO- 氧饱和度仪(Pronto®SpHb®,Masimo 公司)和常用的、侵入性的即时护理、自动化分光光度计(HemoCue 201+®,HemoCue 公司)对非侵入性血红蛋白测量的准确性,与实验室血液分析仪(测量全血血红蛋白)在门诊环境下获得的血红蛋白测量值进行比较。
在门诊研究诊所就诊的成年患者接受 3 种方法的全血血红蛋白测量:非侵入性脉搏 CO- 氧饱和度仪(SpHb®)、即时护理手指采血样本和实验室血液分析仪(参考设备)的静脉血样。计算 SpHb®和 HemoCue 201+®与实验室血液分析仪值相比的偏差和标准差(SD),并生成 Bland-Altman 图。
评估了 152 例样本(平均年龄 46 岁,女性占 69%)。与参考方法相比,SpHb®的偏差±SD 为-0.5±1.0 g/dL,HemoCue 201+®为 0.3±1.0 g/dL。评估测试方法与参考方法一致性的 Bland-Altman 图显示,SpHb®的一致性界限为-2.5 至 1.5 g/dL,HemoCue 201+®为-1.7 至 2.3 g/dL。4 例患者在 2 次尝试后无法获得非侵入性测量值(2.5%的失败率),而所有患者均获得了 HemoCue 201+®的测量值。
非侵入性 SpHb®测试的偏差和 SD 与 HemoCue 201+®相似。由于 SpHb®测量是非侵入性的,它可能为患者和提供者提供额外的益处。