Suppr超能文献

存在噪声时计算机心电图幅度测量的稳定性。CSE工作组。

Stability of computer ECG amplitude measurements in the presence of noise. The CSE Working Party.

作者信息

Zywietz C, Willems J L, Arnaud P, van Bemmel J H, Degani R, Macfarlane P W

机构信息

Medical School Hannover, West Germany.

出版信息

Comput Biomed Res. 1990 Feb;23(1):10-31. doi: 10.1016/0010-4809(90)90003-u.

Abstract

An important feature of an ECG analysis program is its ability to provide reliable measurements under various operating conditions, e.g., on noise-free and noisy ECGs. Therefore, within the European cooperative project "Common Standards For Quantitative Electrocardiography" (CSE), the accuracy and stability of ECG measurements obtained by several computer programs has been compared. To investigate the stability of measurements two sets of 10 ECGs with and without seven different high- and low-frequency types of noise--altogether 160 electrocardiograms and 160 vectorcardiograms--have been analyzed by eight electrocardiographic and five vectorcardiographic computer programs. The stability of measurement was tested with respect to results obtained for the noise-free recordings. In a previous paper, the influence of noise on wave boundary recognition has been reported. In the present paper, the effect of noise on amplitude measurements and on problems of waveform definitions within the QRS complex are described. The results indicate that programs analyzing an averaged beat exhibit less variability than programs which measure every complex or a selected beat. Comparability and stability of measurements could be improved if a standardized procedure for amplitude references were to be introduced. In addition, the stability of QRS waveform labelling could be improved if waveforms' minimum amplitude and duration were to be validated against the noise level which itself should be determined by a standardized procedure.

摘要

心电图分析程序的一个重要特性是其在各种操作条件下提供可靠测量结果的能力,例如在无噪声和有噪声的心电图上。因此,在欧洲合作项目“定量心电图通用标准”(CSE)中,已对多个计算机程序获得的心电图测量的准确性和稳定性进行了比较。为了研究测量的稳定性,八套心电图和五套向量心电图计算机程序分析了两组各10份带有和不带有七种不同高频和低频噪声类型的心电图——总共160份心电图和160份向量心电图。测量的稳定性是根据无噪声记录所获得的结果进行测试的。在之前的一篇论文中,已报道了噪声对波边界识别的影响。在本文中,描述了噪声对幅度测量以及QRS复合波内波形定义问题的影响。结果表明,分析平均心搏的程序比测量每个复合波或选定心搏的程序表现出更小的变异性。如果引入幅度参考的标准化程序,测量的可比性和稳定性可以得到提高。此外,如果根据本身应由标准化程序确定的噪声水平对波形的最小幅度和持续时间进行验证,QRS波形标记的稳定性可以得到提高。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验