To analyze corneal thickness (CT) by anterior segment optical coherence tomography and ultrasound pachymetry in precut endothelial keratoplasty (EK) donor corneas at the eye bank.
Thirty-six corneas were analyzed. All corneas were dissected to create lamellar grafts for EK using the standard eye bank protocol. We measured central and peripheral (3 mm from center) CT by ultrasound after lamellar tissue processing and by optical coherence tomography after the tissue was transferred to a viewing chamber. We performed paired t tests and Pearson correlation analyses to compare ultrasound and optical coherence tomography measurements.
Central CT measurements by optical coherence tomography versus ultrasound in donor corneas were not statistically different (mean: 169 vs. 177 µm, respectively, P = 0.09). There was a strong correlation between measurements of central CT (r = 0.73, P < 0.0001). There was a statistically significant difference in mean peripheral CT measurements by optical coherence tomography and ultrasound (mean: 190 vs. 236 µm, respectively, P < 0.0001); at peripheral locations, measurements correlated moderately (r = 0.52, P = 0.002).
Optical coherence tomography measurements of EK-prepared tissue averaged 8 μm thinner than the ultrasound measurements, similar to clinical measurements of central CT. Given the growing interest in the effect of EK graft thickness on patient outcomes, it is important to understand the differences in measurements obtained by different devices used to assess donor corneas.
在眼库中分析经预处理的内皮角膜移植(EK)供体角膜的眼前节光学相干断层扫描和超声测厚仪测量的角膜厚度(CT)。
对 36 个角膜进行了分析。所有角膜均按照标准眼库方案解剖,以创建用于 EK 的板层移植物。在组织处理后和将组织转移到观察室后,我们通过超声和光学相干断层扫描分别测量中央和周边(距中心 3mm)CT。我们进行了配对 t 检验和 Pearson 相关分析,以比较超声和光学相干断层扫描的测量值。
光学相干断层扫描与超声在供体角膜中央 CT 测量值之间无统计学差异(平均值分别为 169µm 和 177µm,P=0.09)。中央 CT 测量值之间存在很强的相关性(r=0.73,P<0.0001)。光学相干断层扫描和超声测量的外周 CT 平均值差异具有统计学意义(平均值分别为 190µm 和 236µm,P<0.0001);在外周位置,测量值中度相关(r=0.52,P=0.002)。
与超声测量值相比,EK 准备组织的光学相干断层扫描测量值平均薄 8µm,与中央 CT 的临床测量值相似。鉴于人们对 EK 移植物厚度对患者结果的影响越来越感兴趣,了解用于评估供体角膜的不同设备获得的测量值之间的差异非常重要。