评估 38 种常见化妆品在 3 特斯拉磁共振成像中的伪影。

Evaluation of MRI artifacts at 3 Tesla for 38 commonly used cosmetics.

机构信息

Biology Department, Loyola Marymount University, Los Angeles, CA, USA.

出版信息

Magn Reson Imaging. 2013 Jun;31(5):778-82. doi: 10.1016/j.mri.2012.11.002. Epub 2013 Jan 3.

Abstract

PURPOSE

To evaluate MRI artifacts at 3-Tesla for 38 commonly used cosmetics.

MATERIALS AND METHODS

Thirty-eight cosmetics (16, nail polishes; 5, eyeliners; 3, mascaras; 10, eye shadows; 1, lip gloss; 1, body lotion; 1, body glitter, and 1, hair loss concealer) underwent evaluation for MRI artifacts at 3-Tesla. The cosmetics were applied a copper-sulfate-filled, phantom and initially assessed using a "screening" gradient echo (GRE) pulse sequence. Of the 38 different cosmetics, 14 (37%) exhibited artifacts. For these 14 cosmetics, additional characterization of artifacts was performed using a GRE pulse sequence. A qualitative scale was applied to characterize the artifact size.

RESULTS

Artifacts were observed, as follows: 2, nail polishes; 5, eyeliners; 3, mascaras; 3, eye shadows; 1, hair loss concealer. Artifact size ranged from small (eye shadow) to very large (hair loss concealer) and tended to be associated with the presence of iron oxide or other metal-based ingredient.

CONCLUSIONS

Commonly used cosmetics caused artifacts that may create issues if the area of interest is the same as where the cosmetic was applied or if its presence was unknown, thus, potentially causing it to be construed as pathology. Therefore, these findings have important implications for patients referred for MRI examinations.

摘要

目的

评估 38 种常见化妆品在 3T 磁共振成像(MRI)中的磁体伪影。

材料与方法

将 38 种化妆品(16 种指甲油、5 种眼线笔、3 种睫毛膏、10 种眼影、1 种唇彩、1 种身体乳液、1 种身体闪粉和 1 种脱发遮瑕膏)应用于充满硫酸铜的 Phantom 进行评估,最初使用“筛查”梯度回波(GRE)脉冲序列进行评估。在这 38 种不同的化妆品中,有 14 种(37%)出现了伪影。对于这 14 种化妆品,使用 GRE 脉冲序列对伪影进行了进一步的特征描述。应用定性量表来描述伪影的大小。

结果

观察到以下 14 种化妆品出现了伪影:2 种指甲油、5 种眼线笔、3 种睫毛膏、3 种眼影、1 种脱发遮瑕膏。伪影的大小从很小(眼影)到非常大(脱发遮瑕膏)不等,且往往与氧化铁或其他金属基成分的存在有关。

结论

常用的化妆品会产生伪影,如果感兴趣的区域与涂抹化妆品的区域相同,或者如果不知道其存在,可能会导致伪影被误认为是病变,从而产生问题。因此,这些发现对接受 MRI 检查的患者具有重要意义。