Bhaya Maneesha, Mutluer Ferit Onur, Mahan Edward, Mahan Luke, Hsiung Ming C, Yin Wei-Hsian, Wei Jeng, Tsai Shen-Kou, Zhao Guang-Yu, Yin Wei-Hsian, Pradhan Manish, Beniwal Rajesh, Joshi Deepak, Nabavizadeh Fatemeh, Singh Amitoj, Nanda Navin C
Division of Cardiovascular Disease, University of Alabama at Birmingham, Birmingham, AL 35249, USA.
Echocardiography. 2013 Mar;30(3):345-53. doi: 10.1111/echo.12106. Epub 2013 Jan 22.
This study assessed the ability of live/real time three-dimensional transesophageal echocardiography (3DTEE) in measuring (1) atrial septal defect (ASD) maximum dimension, area, and adjacent rim size, (2) ASD occluder left and right atrial disk size, (3) length of contact between the left atrial (LA) disk and the aorta, and in (4) assessing device related complications such as residual shunt, device embolization, and device encroachment upon adjacent cardiac structures.
3DTEE images acquired during percutaneous ASD closure by the Amplatzer Septal Occluder in 15 adult patients were retrospectively analyzed. Offline analysis was done using both the Philips 5500 ultrasound system and Philips QLAB software. 3D color flow Doppler images were used to assess residual ASD shunting.
The Philips 5500 and Philips QLAB measurements correlated well for ASD maximum dimension and area measurements. The Philips QLAB 3DTEE disk size measurements also correlated well with the manufacturer obtained sizes. The aortic rim was deficient in 7 of the 15 patients, and the mean ASD occluder device size was 4 mm greater than the mean ASD maximum dimension. The LA occluder disk was in contact with the aorta throughout the cardiac cycle in 12 of the 15 patients, and the LA occluder disk size correlated significantly with the contact length with the aorta.
Most of the patients demonstrated contact between the LA occluder disk and the aorta throughout the cardiac cycle. 3DTEE may be useful in identifying patients at greater risk for aortic erosion.
本研究评估实时三维经食管超声心动图(3DTEE)测量(1)房间隔缺损(ASD)最大直径、面积及相邻边缘大小,(2)ASD封堵器左、右心房盘片大小,(3)左心房(LA)盘片与主动脉之间的接触长度,以及(4)评估与器械相关的并发症,如残余分流、器械栓塞和器械侵犯相邻心脏结构的能力。
回顾性分析15例成年患者在经皮ASD封堵术中使用Amplatzer房间隔封堵器时获取的3DTEE图像。使用飞利浦5500超声系统和飞利浦QLAB软件进行离线分析。采用三维彩色多普勒图像评估ASD残余分流情况。
飞利浦5500和飞利浦QLAB测量的ASD最大直径和面积相关性良好。飞利浦QLAB 3DTEE测量的盘片大小与制造商提供的尺寸也具有良好的相关性。15例患者中有7例主动脉边缘不足,ASD封堵器平均尺寸比ASD平均最大直径大4 mm。15例患者中有12例在整个心动周期中LA封堵器盘片与主动脉接触,LA封堵器盘片大小与和主动脉的接触长度显著相关。
大多数患者在整个心动周期中显示LA封堵器盘片与主动脉接触。3DTEE可能有助于识别主动脉侵蚀风险较高的患者。