Assessment of intraocular pressure in vitrectomized gas-containing eyes. A clinical and manometric comparison of the Tono-Pen to the pneumotonometer.

作者信息

Lim J I, Blair N P, Higginbotham E J, Farber M D, Shaw W E, Garretson B R

机构信息

Department of Ophthalmology, Eye and Ear Infirmary, University of Illinois, Chicago College of Medicine 60612.

出版信息

Arch Ophthalmol. 1990 May;108(5):684-8. doi: 10.1001/archopht.1990.01070070070037.

Abstract

Elevation of intraocular pressure (IOP) in gas-containing eyes occurs not infrequently, and applanation tonometry in eyes with epithelial irregularities is not accurate. The pneumotonometer and TonoPen (Oculab, La Jolla, Calif) are alternative tonometers for use in these eyes, yet the accuracy of determining IOP with the Tono-Pen has not been determined manometrically. We performed a clinical and manometric study comparing the Tono-Pen with the pneumotonometer in gasfilled vitrectomized eyes. In our clinical study, we measured IOP in 50 eyes of 50 patients with corneal irregularities following vitrectomy and air/gas-fluid exchange. Tono-Pen measurements were highly correlated with those of the pneumotonometer. However, there was a mean difference of 1.4 mm Hg between Tono-Pen and pneumotonometer IOP readings. A larger percentage of Tono-Pen readings were lower than those of the pneumotonometer as IOP levels increased. To assess the accuracy of these tonometers, we compared manometric readings via an indwelling catheter with Tono-Pen and pneumotonometer IOP readings in 11 eye bank eyes following a lensectomy, vitrectomy, and air-fluid exchange. Overall, Tono-Pen and pneumotonometer readings were highly correlated with those of the manometer. However, both machines showed a significant underestimation of IOP at pressures greater than or equal to 30 mm Hg; pneumotonometer and Tono-Pen readings averaged 10.2 and 12.1 mm Hg lower, respectively, than those of the manometer. We conclude that both the pneumotonometer and the Tono-Pen underestimate IOP at pressures greater than or equal to 30 mm Hg.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验