Maudsley A A, Twieg D B, Sappey-Marinier D, Hubesch B, Hugg J W, Matson G B, Weiner M W
Department of Radiology, University of California, San Francisco.
Magn Reson Med. 1990 May;14(2):415-22. doi: 10.1002/mrm.1910140227.
Spectroscopic imaging of phosphorus metabolites in the human brain has been carried out with two data acquisition methods: by observation of the free induction decay (FID) signal and by a short spin echo sequence. The resultant spectral images and spatially resolved spectra are compared. Spin echo observation is found to provide spectra of superior quality, and by suitably selecting the sequence timing, no significant increase in T2 losses, as compared with the FID method, is encountered. 31P images with approximately 3.5 cm spatial resolution are obtained within times of 37 min at 2.0 T field strength.
通过观察自由感应衰减(FID)信号和短自旋回波序列。对所得的光谱图像和空间分辨光谱进行了比较。发现自旋回波观察可提供更高质量的光谱,并且通过适当选择序列时间,与FID方法相比,不会出现T2损失的显著增加。在2.0 T场强下,在37分钟内获得了空间分辨率约为3.5 cm的31P图像。