[使用Reflotron系统测量毛细血管血中的胆固醇。与静脉血清中胆固醇参考测量值进行的约1000次比较的结果]

[Measurement of cholesterol in capillary blood using the Reflotron system. Results from approximately 1,000 comparisons with reference measurements of cholesterol in venous serum].

作者信息

Gerdes L U, Bak A M, Faergeman O

机构信息

Arhus Amtssygehus, medicinsk afdeling I.

出版信息

Ugeskr Laeger. 1990 Jun 11;152(24):1739-43.

PMID:2360292
Abstract

The results of 969 measurements of total cholesterol in capillary blood from fingerstick, performed with the Reflotron-system in the field, were compared with measurements of total cholesterol in venous serum. The comparison allows an evaluation of the combined effect of differences in sample material, working place, stability of methods and staff training. The mean values for all measurement were nearly identical with the two methods, and approximately 95% of the Reflotron-measurements were within an interval of +/- 0.7 mmol/l around the estimated true value. However, values below approximately 6 mmol/l were systematically underestimated, and values above approximately 7 mmol/l were overestimated with the Reflotron-system. The coefficient of variation of the system is below 5%, but both this precision and the accuracy appear to be unstable. Evaluated as a tool in screening for hypercholesterolemia, measurements with the system resulted in a modest extent of erroneous classification of subjects, but the positive diagnostic predictive value of the statement "cholesterol above 7 mmol/l" is only approximately 76%. Simulations of likely variations in the accuracy and precision of the method show a significant influence on the extent of errors. It is recommended that measurements with the system are routinely controlled by measuring an appropriate control material, at the start of each run and after every 30 samples.

摘要

将使用Reflotron系统在现场对969份手指采血的毛细血管血总胆固醇进行的测量结果,与静脉血清总胆固醇的测量结果进行了比较。该比较有助于评估样本材料、工作地点、方法稳定性和人员培训差异的综合影响。两种方法的所有测量均值几乎相同,并且Reflotron测量值中约95%在估计真实值周围±0.7 mmol/l的区间内。然而,Reflotron系统对低于约6 mmol/l的值存在系统性低估,对高于约7 mmol/l的值存在高估。该系统的变异系数低于5%,但这种精密度和准确度似乎都不稳定。作为高胆固醇血症筛查工具进行评估时,该系统测量导致对受试者的错误分类程度适中,但“胆固醇高于7 mmol/l”这一表述的阳性诊断预测值仅约为76%。对该方法准确度和精密度可能变化的模拟显示,其对误差程度有显著影响。建议在每次运行开始时以及每30个样本后,通过测量适当的对照材料对该系统的测量进行常规控制。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验